background image

6  Logger Function

The Gas Detector can record up to 22 hours of detection data. The recorded data can be written 
to a personal computer using optional components that can be purchased separately. (See 10.

 

Consumable Parts and Options.)

Note

●  Maximum gas concentration is recorded every 10 s after logging begins.

●  New log data is stored in the Gas Detector by overwriting old data.

●  The logger function terminates when the Gas Detector is turned OFF.

●  Date settings are not supported.

●  Refer to the XPS-7 Log Data Download Software in the XPS-7L Instruction Manual for the 

procedure used to write log data to a personal computer.

Press the DOWN switch (▼) once in RUN mode. 

   LOG will be displayed on the lower right corner of the LCD display 

and the log start time can be entered.

Press the SPECIAL COMMAND switch and the digit that 

will change starts to flash. Use the UP/DOWN and SPECIAL 
COMMAND switches to set the log start time.

Example:   Enter the numbers shown in the display on the right  to 

set the log start time at 1:20 pm.

After the time has been set, press and hold the ZERO switch. 

MNT1 (maintenance mode 1) will be displayed on the lower left 

corner of the LCD display and log recording will begin.

To end log recording, press the BUZZER STOP switch while LOG 

is displayed on the lower right corner of the LCD display. MNT1 
will no longer be displayed and log recording will stop.

9

8

9

5-3

 

Turning the Gas Detector ON

Press and hold the power switch. The buzzer sounds, the green power lamp flashes, and 

self-diagnosis begins. The diagnostic procedure lasts about 30 seconds.

When self-diagnosis is completed, zero adjustment is performed automatically, the green 

power lamp lights, and the unit is ready to detect gas.

CAUTION

●  The Gas Detector must be turned ON in a clean-air environment. Do not allow the 

Detector to draw in gas while it is warming up. Otherwise, zero adjustment will not 

be performed properly and normal detection will not be possible.

●  Wait at least 1 s before turning the Gas Detector back ON after a power failure.

5-4

 

Detection

Place the gas sampling pipe near the target detection area. When the unit detects gas, the 
concentration of the detected gas is displayed. If the detected gas concentration level reaches the 
alarm point, the alarm lamp lights and the buzzer sounds.

CAUTION

Do not allow water or other liquids to enter the Gas Detector. Any liquid will cause the 

Gas Detector to fail.

5-5

 

Ending Detection

When detection is completed, place the Gas Detector in a clean-air environment and allow it to 
draw in clean air until the concentration indicator reads zero. Press and hold the power switch until 
the Detector turns OFF.

CAUTION

After an adsorbent gas has been detected, replace the filter with a new one 

before attempting to detect another gas. Normal detection is not possible with a 

contaminated filter.

Содержание XPS-7

Страница 1: ...his manual Instruction manual No XPS 7T 10 Semiconductor Plant Gas Detector XPS 7 Instruction Manual Store this Instruction manual in a convenient location and consult it whenever necessary Operate th...

Страница 2: ...ing Procedure 6 5 1 Inserting Batteries 6 5 2 Installing the Sensor Unit 7 5 3 Turning the Gas Detector ON 8 5 4 Detection 8 5 5 Ending Detection 8 6 Logger Function 9 7 Replacing the Filter 10 8 Dail...

Страница 3: ...ath WARNING Indicates a potentially hazardous situation that if not avoided could re sult in serious injury or death CAUTION Indicates a potentially hazardous situation that if not avoided could re su...

Страница 4: ...es FS Alarm setting 1st level 2nd level XDS 7NH NH3 Ammonia 100 ppm 12 ppm 25 ppm XDS 7SH SiH4 Silane 25 ppm 2 5 ppm 5 ppm XDS 7DC SiH2Cl2 Dichlorosilane 25 ppm 2 5 ppm 5 ppm XDS 7AH AsH3 Arsine 250 p...

Страница 5: ...to open it and access the key switches Buzzer speaker Gas inlet Sampled gas aspiration inlet 6 dia Attaches to the gas inlet tube Gas outlet Exhaust outlet for sampled gas 6 dia Adapter connector Con...

Страница 6: ...ecial command switch Used with the logger function UP DOWN switches Used to set the start time and other logger function settings Key Switches LCD Display Log indicator Lit when the log function is op...

Страница 7: ...is attached make sure that it will not pull out Removing Press in the release ring on the coupling and simultaneously pull out the gas inlet tube Attaching Removing Gas Inlet Tube Filter case Contains...

Страница 8: ...AA alkaline batteries provided and make sure they are inserted according to the polarity indicated in the battery compartment Close the battery cover WARNING Replace all four batteries at the same tim...

Страница 9: ...power to the sensor unit A sensor unit that does not have power to it cannot detect gases and may cause the sensor trouble indicator to light A sensor unit installed in the Gas Detector is powered eve...

Страница 10: ...after a power failure 5 4 Detection Place the gas sampling pipe near the target detection area When the unit detects gas the concentration of the detected gas is displayed If the detected gas concentr...

Страница 11: ...e log data to a personal computer Press the DOWN switch once in RUN mode LOG will be displayed on the lower right corner of the LCD display and the log start time can be entered Press the SPECIAL COMM...

Страница 12: ...re factory adjustments that are not available for user application 7 Replacing the Filter Twist the lter holder with detection probe to open it Use a pair of tweezers to remove the O ring and filter I...

Страница 13: ...age 6 The batteries are low Replace the batteries The sensor unit is not installed Install a sensor unit Page 7 Abnormal drop in the ow rate The lter is clogged Replace the lter with a new one Page 10...

Страница 14: ...After examination of allegedly defective product return to New Cosmos with freight prepaid should the product fail to conform to this warranty customer s only remedy and New Cosmos s only obligation s...

Страница 15: ...2 hours of continuous operation at 20 with alkaline batteries and no alarms Operating temperature range 0 to 40 Dimensions W 62 x H 150 x D 128 mm projected portions excluded Mass Approx 1 3 kg Approv...

Страница 16: ...Distributor Manufacturer New Cosmos Electric Co Ltd 2 5 4 Mitsuya naka Yodogawa ku Osaka 532 0036 Japan URL http www newcosmos global com XPS 7 09 T XPS 7T 10 2008 100...

Отзывы: