Chapter 3 - Installation - 37
Cables and Tubing’s installation sequence
1. Connect the power cable of the Quark PFT to the trolley
2. Connect the power cable of the PC and the printer to the trolley
3. Connect the RS232 cable from the Quark PFT (PC RS232) to the PC (COM1 or
COM2). If the PC does not have a RS232 port, please use the USB-RS232 adapter
(A 388 410 001), available as an accessory.
4. Connect the DLCO cylinder to the Quark PFT (DLCO) and adjust its pressure
between 5 and 6 bar.
5. Connect the O2 cylinder to the Quark PFT (O
2
) and adjust its pressure between 5
and 6 bar.
6. Connect the O
2
-CO
2
calibration cylinder to the Quark PFT (CAL) and adjust its
pressure between 5 and 6 bar.
7. Connect the Flowmeter to the front panel of the Quark
Assemble the breathing valve
Содержание Quark PFT
Страница 11: ...Getting started ...
Страница 27: ...Measurements ...
Страница 32: ...32 Quark PFT User Manual ...
Страница 33: ...Installation ...
Страница 42: ...42 Quark PFT User Manual ...
Страница 43: ...Calibration ...
Страница 50: ...50 Quark PFT User Manual ...
Страница 51: ...Database Management ...
Страница 57: ...Spirometry ...
Страница 72: ...72 Quark PFT User Manual ...
Страница 73: ...Lung Volumes ...
Страница 78: ...78 Quark PFT User Manual ...
Страница 79: ...P0 1 Respiratory drive ...
Страница 82: ...82 Quark PFT User Manual ...
Страница 83: ...Lung diffusing capacity ...
Страница 88: ...88 Quark PFT User Manual ...
Страница 89: ...MIP MEP option ...
Страница 91: ...Pulse oximetry option ...
Страница 93: ...Exercise testing ...
Страница 116: ...116 Quark PFT User Manual ...
Страница 117: ...Resting Metabolic Rate Test ...
Страница 123: ...Chapter 13 Resting Metabolic Rate Test 123 ...
Страница 124: ...124 Quark PFT User Manual ...
Страница 125: ...Sub maximal Exercise Testing ...
Страница 131: ...System maintenance ...
Страница 134: ...134 Quark PFT User Manual ...
Страница 137: ...Appendix ...
Страница 148: ...148 Quark PFT User Manual ...