Chapter 13 - Resting Metabolic Rate Test - 119
Recommendations
1. Since the ventilation is very low (normally <10 litres/min), the turbine calibration
has to be performed with very slow manoeuvres (each complete manoeuvre in
about 10-15 seconds), to obtain the best accuracy.
2. Use the following correction for the dead space (V
D
):
- 50 ml for the small mask
- 60 ml for the medium mask
- 70 ml for the large mask