142 - Quark PFT User Manual
Converting factors configuration
You can edit the parameters shown in Control Panel by selecting
Control Panel
from
the
Calibration
menu in the calibration program, then pressing the button by side.
You might configure the following options:
Name
:
identify the parameter
Unit of meas
.:
unit of measurement
Base line and Gain
: factors used to convert the acquired raw data (mV) into the final
format according to
Y=(mV-BL)*Gain
. The value entered for
gain must be multiplied by 1000 (for Gain=1, enter 1000).
Precision
:
the number of decimals shown as
0
Содержание Quark PFT
Страница 11: ...Getting started ...
Страница 27: ...Measurements ...
Страница 32: ...32 Quark PFT User Manual ...
Страница 33: ...Installation ...
Страница 42: ...42 Quark PFT User Manual ...
Страница 43: ...Calibration ...
Страница 50: ...50 Quark PFT User Manual ...
Страница 51: ...Database Management ...
Страница 57: ...Spirometry ...
Страница 72: ...72 Quark PFT User Manual ...
Страница 73: ...Lung Volumes ...
Страница 78: ...78 Quark PFT User Manual ...
Страница 79: ...P0 1 Respiratory drive ...
Страница 82: ...82 Quark PFT User Manual ...
Страница 83: ...Lung diffusing capacity ...
Страница 88: ...88 Quark PFT User Manual ...
Страница 89: ...MIP MEP option ...
Страница 91: ...Pulse oximetry option ...
Страница 93: ...Exercise testing ...
Страница 116: ...116 Quark PFT User Manual ...
Страница 117: ...Resting Metabolic Rate Test ...
Страница 123: ...Chapter 13 Resting Metabolic Rate Test 123 ...
Страница 124: ...124 Quark PFT User Manual ...
Страница 125: ...Sub maximal Exercise Testing ...
Страница 131: ...System maintenance ...
Страница 134: ...134 Quark PFT User Manual ...
Страница 137: ...Appendix ...
Страница 148: ...148 Quark PFT User Manual ...