Instruction Manual | 7
Model Volume
Range (μL)
Counter
Readings
Set
Volume
Increment
8-channel and 12-channel Pipettors
8 - 10
12 - 10
0
3
5
3.5 μL
0.02 μL
8 - 50
12 - 50
0
6
5
6.5 μL
0.10 μL
8 - 200
12 - 200
0
8
5
85.0 μL
0.20 μL
8 - 300
12 - 300
2
5
0
250.0 μL
1.0 μL
To attain the maximum accuracy, the set volume must be
approached from a higher value.
If the desired value is lower than the previous value, adjust
the value down to a one-third turn above the required
setting. If the desired value is higher than the previous
value, adjust the value to a one-third turn above the
required value. The setting should then be performed in
the following manner:
• Slowly turn the pipetting pushbutton or the black
adjustment knob to the required volume.
• If the knob is accidentally turned too far below the
required value, the process must be repeated.
• Set the locking ring into the upper position following
the volume adjustment, thus locking the knob and
preventing accidental volume change.