Instruction Manual | 3
5,000 and 10,000 μL Pipettors
The length of tip ejector is adjusted by screwing in or
screwing out the tip ejector arbor with a small screwdriver.
Turn the screwdriver counterclockwise to increase the length
of the tip ejector, and clockwise to reduce the length of the
tip ejector. The ejector can be adjusted by up to 5 mm.
If the ejector adjustment described above is not sufficient
or the diameter of the ejector opening is too large to eject
the tip, it will be necessary to put the ejector cap “M” onto
the ejector.
2 and 10 μL Pipettors
The 2 μL to 10 μL pipettors may require the use of an
ejector cap “M” to efficiently eject certain brands of pipet
tips. Simply place the cap (supplied with the pipettor)
on the bottom of the pipettor shaft and slide the cap
upwards until it surrounds the bottom of the tip ejector.
L
M
M
5,000 μL: L + 6 m
m
10
,000 μL: L + 10 m
m