Instruction Manual   |   5

Model 

Volume 

Range 

(μL) 

Cat. No.

Volume 

(μL)

Accuracy 

(%)

Precision 

(%)

Fit to Tips 

Without 

Filter 

(μL)

Single-channel Pipettors (continued)

5,000

AP-5000

Min. 1,000

±0.6

≤0.25

5,000

2,500

±0.6

≤0.20

Max. 5,000

±0.5

≤0.15

10,000 AP-10ML

Min. 1,000

±2.5

≤0.60

10,000

5,000

±0.8

≤0.30

Max. 10,000

±0.5

≤0.20

8-channel and 12-channel Pipettors

8 - 10 

12 - 10

AP-8-10 

AP-12-10

Min.

0.5

±10.0

≤8.0

10

5

±4.0

≤2.0

Max.

10

±2.0

≤1.2

8 - 50 

12 - 50

AP-8-50 

AP-12-50

Min.

5

±4.0

≤2.5

200

25

±3.0

≤1.2

Max.

50

±1.6

≤0.6

8 - 200 

12 - 200

AP-8-200 

AP-12-200

Min.

20

±3.0

≤1.5

200

100

±1.5

≤0.8

Max.

200

±1.0

≤0.6

8 - 300 

12 - 300

AP-8-300 

AP-12-300

Min.

50

±1.6

≤1.5

300

150

±1.2

≤1.0

Max.

300

±1.0

≤0.6

The accuracy and precision are obtained gravimetrically using 

Axygen® pipet tips, performing at least 10 measurements 

of distilled water at a reference temperature of 20°C ± 1°C 

according to EN ISO 8655 standard.

 2.  Setting  the  Volume

The volume of the pipettor is set by turning the pipetting 

pushbutton (A2) or the black adjustment knob (B). 
The volume display consists of three numbers and is 

read from top to bottom. The three numbers indicate the 

volume selected and are colored black and red.
Red numbers represent decimal fractions for microvolume 

pipettors or milliliters for macrovolume pipettors.

Содержание Axygen Axypet

Страница 1: ...A Corning Brand Axygen Axypet Pipettors Single channel 8 channel and 12 channel Pipettors Instruction Manual...

Страница 2: ...3 Aspiration and Dispensing Instructions 8 4 Pre rinsing 9 5 Dense and Viscous Liquids 10 6 Axygen Pipet Tips 10 7 Recommendations 10 8 Recalibration 11 9 Troubleshooting 15 10 Cleaning and Sterilizat...

Страница 3: ...ranges 0 5 to 10 L 5 to 50 L 20 to 200 L 50 to 300 L Model Volume Range L Application 2 10 Measurement and transfer of micro volumes DNA sequencing and enzyme assay applications 20 50 100 200 250 300...

Страница 4: ...lity of cross contamination of the liquid samples To protect the user from contaminated pipet tips the pipettor is fitted with a built in tip ejector D Single channel Pipettor Ejector The ejector leng...

Страница 5: ...to 5 mm If the ejector adjustment described above is not sufficient or the diameter of the ejector opening is too large to eject the tip it will be necessary to put the ejector cap M onto the ejector...

Страница 6: ...volume depends on the quality of pipet tips used The values for accuracy and precision shown in the table below were obtained using Axygen pipet tips Model Volume Range L Cat No Volume L Accuracy Pre...

Страница 7: ...0 1 5 200 100 1 5 0 8 Max 200 1 0 0 6 8 300 12 300 AP 8 300 AP 12 300 Min 50 1 6 1 5 300 150 1 2 1 0 Max 300 1 0 0 6 The accuracy and precision are obtained gravimetrically using Axygen pipet tips pe...

Страница 8: ...L Counter Readings Set Volume Increment Single channel Pipettors 2 1 2 5 1 25 L 0 002 L 10 0 7 5 7 5 L 0 02 L 20 1 2 5 12 5 L 0 02 L 50 0 3 8 38 0 L 0 2 L 100 0 7 5 75 0 L 0 2 L 200 1 2 5 125 0 L 0 2...

Страница 9: ...lue adjust the value down to a one third turn above the required setting If the desired value is higher than the previous value adjust the value to a one third turn above the required value The settin...

Страница 10: ...ct approximately 1 5 mm into the manifold The suspension system of the shafts ensures even and tight sealing of the pipet tips The rocking movement does not have to be performed to seal the pipet tips...

Страница 11: ...ters A replaceable filter L fitted in a seat in the bottom part of the shaft is used in 5 000 L and 10 000 L pipettors This filter prevents the aspirated liquid from entering into the shaft and thus f...

Страница 12: ...ff from the nominal value Correction value 2 x nominal val m weight of the sample density of liquid Check this operation once again and correct if necessary Note the corrected value for further pipett...

Страница 13: ...ns which emit vapors it is recommended to disassemble the shaft and rinse the piston and seal with distilled water after finishing the pipetting operation 8 Recalibration The accuracy and precision of...

Страница 14: ...5 C and stabilized during weighing within 0 5 C Measurements should be conducted using distilled water Balance sensitivity should be suitable for the volume to be measured Recalibration Procedure Set...

Страница 15: ...aspirations weigh each one and calculate the average value of the aspirations Perform 3 aspiration series each series should include the aspirations from all channels weigh each time and calculate th...

Страница 16: ...ll separate 2 Holding the volume setting knob to protect it against rotation insert the calibration key into the cuts of the calibration screw 3 Turn the calibration key clockwise to reduce the aspira...

Страница 17: ...acement of parts should be required only occasionally and should not occur under normal pipettor use Problem Cause Solution Droplets of liquid remain in the pipet tip The pipet tip is emptied too fast...

Страница 18: ...ored Remove the tip ejector Unscrew the shaft nut and inspect the shaft and the piston assembly Replace the damaged parts When reassembling the pipettor the nut should be hand tightened In the single...

Страница 19: ...o lubricant on the sealing elements Remove the tip ejector Unscrew the shaft nut and remove the shaft piston assembly seal and O ring Rinse the removed parts in distilled water and dry thoroughly Ligh...

Страница 20: ...rilization and re tighten after autoclaving Set the locking ring in lower unlocked position prior to sterilization Sterilize the pipettors in autoclave with an initial vacuum and drying cycle Prior to...

Страница 21: ...ters for pipettor models AP 5000 AP 10ML Lubricant for single channel pipettors Color identification rings for single channel pipettors Right hand ejector button Cat No Description AP USD Pipettor sta...

Страница 22: ...hase CORNING DISCLAIMS ALL OTHER WARRANTIES WHETHER EXPRESSED OR IMPLIED INCLUDING ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR OF FITNESS FOR A PARTICULAR PURPOSE Corning s sole obligation shall be t...

Страница 23: ...xclusion or limitation of incidental or consequential damages This warranty gives you specific legal rights You may have other rights which vary from state to state No individual may accept for or on...

Страница 24: ...an t 886 2 2716 0338 EUROPE CSEurope corning com France t 0800 916 882 Germany t 0800 101 1153 The Netherlands t 20 655 79 28 United Kingdom t 0800 376 8660 All Other European Countries t 31 0 206 59...

Отзывы: