124
EM 9250
Operator’s manual
• press the
key
• use the
keys to search
for GAU CAL
• press the
key to perform cali-
bration, proceeding as follows:
1. The machine will display the following
message:
•
CAL on the left display
•
E flashing on the right display
2. Fix the calibration template in cor-
respondence of the threaded hole
located in the shaft assembly bell using
the M8 screw (A, Fig.25) supplied with
the ultrasonic sensor.
3. Use the wheel fixing nut to bring the
template in contact with the shaft as-
sembly bell (fig.25).
4. Press the
key to confirm the
securing of the template.
5. Once confirmed, “CAL rot” is shown
on the display.
6. Slowly turn the template towards the
operator to automatically enable the
clamping brake (fig.26).
7. When template movement is complete,
the message “CAL CLO” will appear on
the display.
8. Slowly lower the wheel guard (fig.
27), the machine will automatically
calibrate the sensor.
If the calibration has been performed
successfully, an acoustic consent signal
is emitted.
On the contrary, the display of the mes-
sage A20 indicates that:
- the position of the calibration template
during the calibration phase is not
correct. Therefore position it correctly,
25
26
27