PoolWarden Manual www.poolwardentraining.com 530-205-4520 25
Step 5 - Service Menu
The Service Menu includes all items that a service technician needs to
service a pool and is a critical next step in setting up PoolWarden. From the
Default Readings Screen simply press the
Back
button to access the main
menu and select the first line “Service” and press the
Enter
button to access
the Service Menu. From here you will be able to manually turn on and off all
relays, calibrate pH and ORP sensors, disable the relays for servicing and
clear all the overfeed timers.
⧫
Relay Manual Mode:
Use the up and down arrow keys to scroll next to
the relay that needs to be put into manual mode. Press the
Enter
button
or the
Right Arrow
to put the selected relay into manual mode. There are
three states that the relay can be in when in manual control:
❑
If the relay is currently ON, it will turn off for the amount of manual
relay time in the relay setup. Once this time is finished the relay will go
back to auto.
❑
If the relay is OFF, and not in manual OFF mode, it will turn ON for the
manual relay time and return to auto.
If the relay is OFF, and not manual mode, it will go back to auto.
⧫
Calibration Pool1 and Pool2:
Select the desired pool to calibrate and
scroll to the item to calibrate. Manual Pool Measurements must be taken
to calibrate each sensor. For best results the pool should be at the desired
values when calibrating. The percentage to the right of the current reading
is the amount of calibration and if it is at 99% that sensor is at the
maximum calibration and may need to be replaced.
❑
pH: Enter the manually measured pH value. A pH sensor can be
cali/- 2 pH units.
❑
ORP: If the sanitizer level is higher than desired and the pH is at the
setpoint then raise the ORP calibration slightly. If the sanitizer level is
lower than desired and the pH is at the setpoint then lower the ORP
calibration slightly. The ORP sensor can be cali/- 200 mV.
ORP sensors are affected by cyanuric acid, pH and other factors and it
may take a few days to get it adjusted.
Relay Manual
Calibration P1
Calibration P2
Disable Relays
Clear Overfeed
TrueDPD PPM Manual
DPD Level 55%
Service Menu
❑
Temperature: Enter the manually measured temperature. A temperature sensor can be calibrated +/-
25 degrees Fahrenheit.
❑
PPM: If the TrueDPD is connected there will be a selection for calibrating the free chlorine. When it is
calibrated, the calibration is calculated on the next measurement cycle, if the PoolWarden power is
cycled before that happens the calibration will be lost.
Disable Relays:
To make sure the chemical feeders do not turn on while servicing a pool. Press
Enter
up
to 4 times to disable from 15 minutes to 60 minutes. Press the
Enter
key again to put the relays back to
auto.
⧫
Clear Overfeed Times:
If an overfeed timer has been reached it will only clear/reset at midnight. There
are times when servicing a pool that you may want to clear the overfeed timers to stop an alarm from
tripping or to have the relay turn back on.
❑
The only way to clear the overfeed timers is to select Clear Overfeed in the “Service” menu or to wait till
midnight when they automatically clear. Note: Cycling power on the PoolWarden will not clear the
overfeed timers.
❑
The Setpoint Overfeed timers are not cleared when selecting the Clear Overfeed Times.
⧫
TrueDPD Manual:
If the TrueDPD is connected, this menu allows for manually controlling all of the
functions of the TrueDPD which is useful to make sure everything is working properly and purging the
DPD.
⧫
DPD Level 55%:
Displays the current DPD level. Press enter to change the value in 10% increments.
PH 1
00:00:00 OFF
ORP 1
00:00:00 OFF
Aux 1
00:00:00 OFF
Aux 2
00:00:00 OFF
PH 2
00:00:00 OFF
ORP 2
00:00:00 OFF
Aux 3
00:00:00 OFF
Aux 4
00:00:00 OFF
Relay Manual
Cal ORP1 650 -16%
Cal pH1 7.7 -5%
Cal Temp1 78
+12%
Clear Cal ORP1
Clear Cal pH1
Clear Cal Temp1
Cal PPM1 3.4 -6%
Clear Cal PPM1
Calibration