17
C561 • C562
C561 • C562
18
Specifications
C561
C562
pH
Range
0...14 pH
0...14 pH
Resolution
0.01 pH
0.01 pH
Accuracy
0.2% ± 1 digit
0.2% ± 1 digit
Calibration
1...3 points
1...3 points
Buffers
9 pre-programmed
9 pre-programmed
Temperature Compensation
0...100°C
0...100°C
ISO-pH
6...8 pH
6...8 pH
Slope
80...120%
80...120%
mV
Range
±1000 mV
±1000 mV
Resolution
1 mV
1 mV
Accuracy
0.2% ± 1 digit
0.2% ± 1 digit
Calibration
1 point
1 point
CONDUCTIVITY
Range
0...1000 mS/cm
0...1000 mS/cm
Resolution
0.01 µS/cm
0.01 µS/cm
Accuracy
1% f.s. of range
1% f.s. of range
Calibration
1 point
1 point
Standards
0.01/0.1/1M KCl
0.01/0.1/1M KCl
Cell Constant
0.1/1/10 cm
-1
± 30%
0.1/1/10 cm
-1
± 30%
Temperature Compensation
0...100°C
0...100°C
Reference Temperature
20° or 25°C
20° or 25°C
Temperature Coefficient
natural waters (EN27888)
natural waters (EN27888)
SALINITY
Range
0...70 ppt
0...70 ppt
Resolution
0.1 ppt
0.1 ppt
TDS
Range
0...100 g/l
0...100 g/l
Resolution
0.1 mg/l
0.1 mg/l
DISSOLVED OXYGEN
Range
0...20 mg/l (0...200%)
Resolution
0.05 mg/l (0.5%)
Accuracy
1% ± 1 digit
Calibration
1 point
Temperature Compensation
0...50°C
Salinity Compensation
0...40 ppt
Air Pressure Compensation
800...1200 hPa
TEMPERATURE
Range
0...100°C
0...100°C
Resolution
0.1°C
0.1°C
Accuracy
0.5°C
0.5°C
Calibration
1 point
1 point
INPUTS
pH/mV
BNC, 10
12
Ω
BNC, 10
12
Ω
Conductivity/Dissolved Oxygen
BNC
BNC
Temperature
2 banana, for Pt1000
2 banana, for Pt1000
DATA-LOGGING
Memory
300 temperature
300 temperature
DISPLAY
LCD
122x32 pixels
122x32 pixels
White Backlight
ü
ü
Languages
EN, NL, FR, DE
EN, NL, FR, DE
POWER SUPPLY
Mains
210...250 VAC, 50/60 Hz
210...250 VAC, 50/60 Hz
Low Voltage
9...15 VDC
9...15 VDC
Batteries
4x1.2 V, NiMH
4x1.2 V, NiMH
DIMENSIONS
WxDxH
10x20x4 cm
10x20x4 cm
WEIGHT
Meter
400 g
400 g