Calibration:
If you have ordered the custom 8 thermocouple kit, then your kit came with two EGT modules. These
modules convert 4 analog thermocouple signals into one digital wire that then plugs into a CAN Bus
port on the Main module. To calibrate your system, simply plug your EGT modules into available CAN
Bus ports on the Main module, record a quick test run, and download the SD card to your PC. The
Main module will sense the presence of the EGT module and will automatically set the calibration for
thermocouples.
If you have ordered a single thermocouple and are using an analog channel to monitor, you will need to
adjust the calibration setting for that analog channel. To configure, open the Calibration/Configuration
window, and click on the “configure” button for the specific channel that you have plugged the
thermocouple into. You can now change the type of the “Type of Sensor” to “EGT (1800 degree range)
from the drop down menu.
Testing:
If the engine is off and cold, the readings should be within 20 degrees Fahrenheit of room temperature.
If the engine is warm, appropriate readings should be seen. If your probes are reading close to 30
degrees, then there is very likely an accidental reversal of wires somewhere. If your probe is reading a
maximum or close to 1800 degrees, then there is an open circuit, and most likely a wire unplugged.
Check all connections between the actual probe and the
D
ATA
M
AXX
module.