46
8.5
Cycle Printout - Full Cycle
STATIM 6000B
SDS2R100
SN 310800A03600
UNIT # :
000
WATER QUALITY
5.2uS / 3.3ppm
CYCLE NUMBER
000043
13:31:12
2020/12/15
Solid Wrapped (S)
134 C/3.5min
WARMING UP
111.7 C 101kPa
0:00
VACUUM DRAW
111.6 C 17kPa
1:11
PRESSURE PULSE
111.9 C 121kPa
1:23
VACUUM DRAW
111.2 C 29kPa
1:56
PRESSURE PULSE
112.5 C 121kPa
2:12
VACUUM DRAW
112.1 C 29kPa
2:48
PRESSURE PULSE
114.2 C 121kPa
3:05
STERILIZING
5:47
135.7 C 315kPa
5:47
Min. steri. Values:
135.3 C 311kPa
Max. steri. Values:
135.9 C 317kPa
VENTING
9:17
135.5 C 314kPa
9:17
DRYING START
11:28
117.2 C 79kPa
CYCLE COMPLETE
21:34
Digital Signature #
9E7726C95F4CDA91
12D2D6DCF5BBC248
B9106A8FC7F49F08
365400FA91D368AF
Model:
STAT
IM
6000B
Serial Number:
310800A03600
Unit Identifier:
Autoclave has
been set up as number 000
Quality of water in reservoir
Time/Date:
1:31 pm
12 December 2020
Cycle clock:
starting at 0:00
Software:
Revision 100
Temp./Press. & Time
printed
at different intervals during the
vacuum draw and pressure pulse
phases
Start Time of Sterilisation:
5:47 (start of ‘D’ phase)
Start Time of Venting:
9:17 (start of ‘E’ phse)
Start Time of Air Drying:
11:28 (start of ‘F’ phase)
Cycle completion time:
21:34
Sterilisation Parameters
Cycle counter:
the number
of cycles that were run on the unit
is 43.
Cycle name and Parameters:
SOLID WRAPPED (S)
for 134°C/3.5 minutes
Digital signature of the unit
Warm up complete:
Start of the vacuum
draw is 1:11