Connecting to a Waste Bottle
To connect the waste bottle to the STAT
IM
B, follow these steps:
1.
Insert the exhaust tube into the fitting on the
back of the unit and pull gently to confirm it is
holding correctly.
2.
Cut the tube to length and slide the other end
into the fitting on the bottle.
3.
Unscrew the lid from the bottle.
4.
Fill the bottle with water to the MIN line and re-
install the lid and copper condenser assembly.
5.
Place the waste bottle below the unit to allow
proper drainage.
IMPORTANT!
Avoid excess sagging in the drain
tube, cut the tube to measure.
IMPORTANT!
Tube should not be kinked,
bent or otherwise obstructed. The connection
point to the waste bottle or central drain must
be lower than the autoclave’s support surface
otherwise the unit may not drain correctly.