33
5.6 Running a Helix Test
To perform a Helix test, you will need a Helix device or test pack. These are NOT supplied with your
STAT
IM
B. To perform the test, follow the instructions provided by the test pack manufacturer.
Generally, the process is as follows:
1.
Open the drawer to insert a Helix test pack.
Position pack at the front on the lower tray.
2.
Close and lock the drawer.
3.
From the home screen, select the
TESTS
icon.
4.
Select the
HELIX TEST
icon.
5.
Press
START NOW
.
6.
Press the
STOP
icon to open the drawer.
7.
Follow the test manufacturer’s instructions to
interpret the test results.
8.
If the unit passes the test, it is ready for use.
If the unit fails, check the test manufacturer’s
instructions and repeat the test. If the second
test fails, contact your service technician.
Presetting your Helix test
To schedule a Helix test before the start of the next working day, use the delayed start function. The process
is the same as described in the section above for the Bowie-Dick test.