34
5.7
Running a Vacuum Test
The vacuum test checks the autoclave’s plumbing system for leaks and should be done on a regular basis
in accordance to your local guidelines. Run this test with the empty trays in the chamber. Test should be
done on a cold chamber. If the chamber is hot, turn OFF the unit (or turn off the stand-by mode) and let the
drawer cool.
1.
From the home screen, select the
TESTS
icon.
2.
To run a vacuum test, press the
V icon
.
3.
Press
START NOW
.
Running a vacuum test can take a minimum of 15 minutes. When the test is complete, the screen will
display a CYCLE COMPLETE message. If the test has failed,
see Section 11. Troubleshooting.
Presetting your vacuum test
To schedule a vacuum test before the start of the next working day, use the delayed start function. The
process is the same as described in the section above for the Bowie-Dick test.