Sterilizers Bravo G4
- TECHNICAL MANUAL
54
ERROR/WARNING-ALARM-HAZARD-SYSTEM ERROR
EN
CODE AND
MESSAGE ON THE
DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT (**)
E009
QUALITY H
2
O BAD
CHANGE FILTERS
CHANGE WATER
H
2
O conductivity
above 60 uS/cm
(automatic filling with
centralized system or
Pure 100/500 - ROW
versions only).
NOTE:
IN THESE
CONDITIONS, THE
DEVICE ALLOWS
STARTING UP TO 5
CONSECUTIVE
CYCLES, THEN IT
LOCKS OUT UNTIL
WATER WITH
PROPER QUALITY
IS DETECTED (<15
μS/cm).
THIS PRECAUTION
IS NECESSARY TO
PREVENT ANY
DAMAGE TO THE
DEVICE.
For connection to
Demineralizer PURE
100 – device check.
(ROW versions only.
Empty the filling tank and replace
the 2 PURE 100 resin cartridges.
(ROW versions only).
Top up the tank, activate the
sterilization cycle to reset the
cycle counter for autoclave lock.
1/2
For connection to
Demineralizer PURE
500 - device check.
(ROW versions only.
Empty the filling tank and
replace the PURE 500 filtering
element kit.
(ROW versions only).
Top up the tank, activate the
sterilization cycle to reset the
cycle counter for autoclave lock.
For connection to
Demineralization
centralized system -
check the quality of the
water supplied by the
system.
Empty the filling tank and select
manual supply mode and fill the
tank with Demineralized water
falling within the limit of 15
μS.
Check and restore the correct
conductivity value of the water
supplied by the centralized
system, select automatic supply,
top up the tank, activate the
sterilization cycle to reset the
cycle counter for autoclave lock.
Conductivity sensor
problem.
Check that the water in the tank
has an appropriate conductivity
using an external sensor. Empty
the filling tank, replace the
conductivity sensor and fill it
with demineralized water falling
within the limit of 15 μS. Activate
the sterilization cycle to reset
the cycle counter for autoclave
lock.
E010
DOOR OPEN
CLOSE DOOR
Door open.
The door is not
properly closed.
Properly train the operator on the
correct ways to close and open
the door.
1
Door closing activation
failure - check ajar
door microswitch.
Check the correct operation of the
microswitch by means of the
SERVICE/COMPONENT
TEST/DOOR menu.
Replace the microswitch - see the
technical sheet DOOR CLOSING
UNIT.
Содержание SciCan BRAVO G4
Страница 2: ......
Страница 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Страница 12: ...Sterilizers Bravo G4 TECHNICAL MANUAL 10 GUIDELINES SAFETY TEST VALIDATION IF REQUIRED EN 3 2 CONFORMITY REPORT...
Страница 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Страница 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Страница 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Страница 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Страница 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Страница 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Страница 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Страница 91: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN TECHNICAL SHEETS 89 1 Steam generator complete with heat insulation...
Страница 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...
Страница 116: ......