Sterilizers Bravo G4
- TECHNICAL MANUAL
EN
TECHNICAL MAINTENANCE
109
8.2. SCHEDULED MAINTENANCE MESSAGES
The sterilizer periodically displays warning messages relevant to "routine" maintenance operations, required to
ensure the proper operation of the device.
Press DONE to confirm that the required operation has been
made.
The relevant cycles counter will be reset.
Press REMIND button to postpone the operation.
The warning message will reappear the next time the sterilizer
is used.
The cycle counter, related to SCHEDULED MAINTENANCE MESSAGES display, can reset in advance
using the SERVICE RESET menu (SQ34) from SERVICE mode.
8.3. DESCRIPTION OF MAINTENANCE INTERVENTIONS
8.3.1. DOOR GASKET AND THE INTERNAL PART OF THE DOOR CLEANING
To eliminate any traces of limescale, clean the gasket of the chamber and the door porthole with a clean cotton cloth
soaked in a soft solution of water and vinegar (or a similar product, checking the contents on the label before using).
Dry the surfaces and remove any residues before using the device.
8.3.2. STERILIZATION CHAMBER AND ACCESSORIES CLEANING
Clean the sterilization chamber, support and trays (and internal surfaces in general) with a clean cotton cloth soaked
in water and, possibly, the addition of a small amount of neutral detergent.
Carefully rinse with distilled water, taking care not to leave any type of residue in the chamber or on accessories.
Do not use pointed or sharp tools to remove scale from the sterilization chamber.
Should there be evident deposits, immediately check the quality of the distilled water used (see technical
characteristics appendix).
8.3.3. EXTERNAL SURFACE CLEANING
Clean the external surfaces using appropriate product (i.e. Ethyl alcohol, 50% diluted with water).
Apply product with a soaked cloth, then dry.
Do not spray or vaporise any product directly on device surfaces.
Inflammable liquid.
8.3.4. CHAMBER FILTER CLEANING
While using the sterilizer, various residues accumulate in the filter and with time obstruct the lower drain duct. To
clean the filter, open the sterilizer door and remove the cap using a coin or another suitable tool.
Loosen the union that contains the filter. Remove the filter from its support and thoroughly clean it under a running
water jet, if necessary using a sharp tool to remove any large foreign bodies. It is possible to use a compressed air
jet. If it is impossible to recover the filter, replace it with a new one.
Refit everything operating in reverse order and making sure to screw the union, positioning the drain holes at the
level of the boiler wall.
Properly fit the filter in its housing, a partial fitting may damage the component.
Содержание SciCan BRAVO G4
Страница 2: ......
Страница 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Страница 12: ...Sterilizers Bravo G4 TECHNICAL MANUAL 10 GUIDELINES SAFETY TEST VALIDATION IF REQUIRED EN 3 2 CONFORMITY REPORT...
Страница 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Страница 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Страница 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Страница 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Страница 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Страница 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Страница 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Страница 91: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN TECHNICAL SHEETS 89 1 Steam generator complete with heat insulation...
Страница 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...
Страница 116: ......