Sterilizers Bravo G4
- TECHNICAL MANUAL
EN
SERVICE MENU
49
SQ32 – WATER LOADING TIMES
The time settings for the water filling can be set automatically or manually.
AUTOMATIC SETTING:
With empty tank, click on START to activate the automatic water filling.
The device detects the time necessary to reach the minimum level (alarm
time) and the time required to reach the maximum level. At the end of the
procedure, the display shows the new values.
Press ENTER and reconfirm the new values.
MANUAL SETTING:
The technician must manually set the filling time and the alarm time, by
means of the counters.
Press ENTER to confirm.
Pre-set times are as follows:
External Pump Option:
-
Alarm Time: 170 seconds
-
Filling Time: 150 seconds
Pure 100 Option: (ROW versions only)
-
Alarm Time: 1000 seconds
-
Filling Time: 500 seconds
Pure 500 Option: (ROW versions only)
-
Alarm Time: 2500 seconds
-
Filling Time: 900 seconds
Ev Aux Option:
-
Alarm Time: 1000 seconds
-
Filling Time: 1000 seconds
SQ33 – SERVICE RESET
It allows resetting the counters of the intervals of the maintenance
messages.
It must be used when a maintenance operation is performed in advance.
SQ34 - BACKUP MESSAGE
Select BACKUP MESSAGE to display a screen where it is possible to
select whether to activate/deactivate backup and overwrite messages.
In particular, the backup message warns the user to run a backup every
250 cycles.
The overwrite message warns the user that the oldest data will be
overwritten.
The values can be modified by:
-
Changing the value (ON/OFF).
-
Pressing ENTER to confirm.
The user may reset the BACKUP and OVERWRITE messages by
downloading the cycles using the NEW option.
Содержание SciCan BRAVO G4
Страница 2: ......
Страница 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Страница 12: ...Sterilizers Bravo G4 TECHNICAL MANUAL 10 GUIDELINES SAFETY TEST VALIDATION IF REQUIRED EN 3 2 CONFORMITY REPORT...
Страница 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Страница 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Страница 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Страница 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Страница 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Страница 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Страница 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Страница 91: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN TECHNICAL SHEETS 89 1 Steam generator complete with heat insulation...
Страница 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...
Страница 116: ......