Sterilizers Bravo G4
- TECHNICAL MANUAL
EN
ERROR/WARNING-ALARM-HAZARD-SYSTEM ERROR
51
6. ERROR/WARNING-ALARM-HAZARD-SYSTEM ERROR
6.1. CODE “E” ERROR/WARNINGS
(**)
1
= OK (warning) - Tap OK on pop-up window.
2
= OK + door unlock / RESET - Keep OK pressed for 3 seconds on the pop-up window and the autoclave will start
the procedure to open the door in full safety. When the RESET key appears in the left area of the LCD, hold it
pressed for 3 seconds.
CODE AND
MESSAGE ON THE
DISPLAY
DESCRIPTION
Possible causes /
Checks
Solution
ERROR
OUTPUT (**)
E000
POWER OUTAGE
CONTACT TECH.
SERVICE
Blackout
A power voltage failure
occurred.
Once the power is back, follow
the indications on the display to
open the door. Reset the alarm
and repeat the sterilization cycle.
2
Autoclave accidentally
turned off with the
main switch.
Turn the autoclave on again and
follow the indications on the
display to open the door. Reset
the alarm and repeat the
sterilization cycle.
Autoclave accidentally
turned off due to the
disconnection of the
power cable from the
power socket.
Reconnect the power cable
properly and follow the indications
on the display to open the door.
Reset the alarm and repeat the
sterilization cycle.
Alarm shown at every
start-up.
Check that the operator has
properly reset the alarm and give
advices on its correct use.
Problem with fuse
holders.
Check the mains 16 A fuse
continuity with tester.
Replace the blown fuse with one
having the same rate. Turn the
autoclave on again and follow the
indications on the display to open
the door.
Check the plug of the
fuse holder.
Make sure that the plug is not
excessively tightened.
Tighten the plug correctly by
slightly loosening it.
Problem with the
power cable - cable
check.
If the cable is damaged, replace
it.
E001
OVERVOLTAGE
CONTACT TECH.
SERVICE
Overvoltage
Supply voltage value
beyond limits - check
supply voltage.
Disconnect the power cable and
check the mains voltage. The
value must be within the rated
voltage /- 10%.
If it exceeds the limits, do not
connect the autoclave and check
the power mains.
1
Excessive voltage
value at transformer
output - check output
voltage.
Disconnect the wiring at
transformer output and check
voltage. The value must not
exceed 34V.
Replace the transformer.
Содержание SciCan BRAVO G4
Страница 2: ......
Страница 4: ...Sterilizers Bravo G4 TECHNICAL MANUAL 2 CONTENTS EN ITALIANO Page left blank intentionally...
Страница 12: ...Sterilizers Bravo G4 TECHNICAL MANUAL 10 GUIDELINES SAFETY TEST VALIDATION IF REQUIRED EN 3 2 CONFORMITY REPORT...
Страница 23: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 21...
Страница 25: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 23...
Страница 27: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 25...
Страница 29: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 27...
Страница 33: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 31...
Страница 35: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 33...
Страница 39: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN FUNCTIONING 37...
Страница 91: ...Sterilizers Bravo G4 TECHNICAL MANUAL EN TECHNICAL SHEETS 89 1 Steam generator complete with heat insulation...
Страница 98: ...Sterilizers Bravo G4 TECHNICAL MANUAL 96 TECHNICAL SHEETS EN 1 PCB 2 Pressure transducer...
Страница 116: ......