Measuring / Loading a standard
Measuring / Loading a standard
68
ColorLite GmbH | sph860/sph900
8
Trouble shooting
The sph860/sph900 is manufactured using the latest
components and state of the art technology.
A Manipulation inside the device by the user is not
needed.
Do not open the device and or the probe head.
Opening the housing can damaged visual and me-
chanically sensitive components.
An unauthorised opening of the device or probe head or
will make all warranty and warranty claims against the
ColorLite GmbH invalid
Problem
Solution
The device cannot be
switched
on.
The
display remains off
Reload the rechargeable bat-
tery
No remission spec-
trum is displayed after
a loading a standard
from the memory.
Colour standards can be stored
with or without the remission
data. To store with remission
spectrum
switch on
in the
“Settings”
window
“Save
spectrum”
Communication
with
the PC is not working.
Check the cable connection.
Check that COM-port settings
are correct. When the device is
connected via USB the port
should be listed. Also if the
Bluetooth driver has been cor-
rectly installed the device serial
no. should be visible.
Check that a valid license key
has been installed in Color-
DaTra (menu: spectrophotome-
ter – license overview)
Warning window:
Check that the calibration
standard is correct and not
contaminated.
Check the probe head opening
for contamination and clean if
necessary with
dry and oil
free
compressed
air.
If this does not help please
contact the
ColorLite GmbH
or your local distributor.
If you have reason to assume that the sph860/sph900 is
not working correctly or results do not correspond to the
device specifications, please contact the ColorLite GmbH
or your local dealer.