Section 4 Operation
1.
With
Password/Reset
highlighted, press
to display:
NOTE
The controller resets to the
User
mode
whenever the unit is turned off. The control
-
ler also resets to the
User
mode whenever the
Start/Status Display is displayed continuously for
10 minutes.
2.
Press and change the number to
444
.
Highlight the desired reset option and press
.
A confirmation message will appear, press
again.
1.
If desired, highlight
Reset
and press
to
display:
Level User
Password 1
Menu
Level Factory
Password 444
Reset
Calibration
Menu
Reset user settings
Reset PID settings
Reset both
Menu
Level User
Password
444
Menu
System - Password/Reset
is used only by a qualified technician. Changing the password enables
controller reset options, the temperature sensor calibration procedure and displays PID values.
3.
Press
to display:
Scroll down to display
PID Tuning
.