4-8
Cole Palmer
Section 4 Operation
1.
With
Alarms
highlighted, press
to
display:
2.
With
Temperature Alarms
highlighted,
press
to display:
1.
With
Audible Alarms
highlighted, press
to display the alarms.
Highlight the desired alarm and press
to
toggle between enable and disable mode.
High Fault
cannot be set below
High Warn
.
High Warn
cannot be set below
Low Warn
.
Low Fault
cannot be set above
High Warn
.
If
Faults
is enabled the alarm will sound when a
fault occurs. If
Warnings
is enabled the alarm
will sound when a warning occurs.
Menu
Temperature Alarms
Audible Alarms
Low Level Warning
Menu
Temperature Alarms
Audible Alarms
Low Level Warning
Menu
High Fault
83.0°C
High Warn
83.0°C
Low Warn
2.0°C
Low Fault
2.0°C
3. Highlight the desired limit and press
. Follow the same procedure used to change a setpoint.
If the Fault temperature is exceeded the unit will shut down and, if enabled, the audible alarm will
sound. If the Warn temperature is exceeded the unit will continue to run and, if enabled, the audible
alarm will sound. In both cases a message will be displayed.
Menu
Faults
Warnings
Press
, or use the
Menu
window, to save
and return to the previous display.
Press
, or use the
Menu
window, to save
and return to the previous display.
Press
, or use the
Menu
window, to save
and return to the previous display.
1.
With
Low Level
highlighted, press
to
toggle the low level warning alarm on/off:
Alarms
is used to view/adjust the high and low temperature alarm limits, to enable/disable the audible
alarms and to configure the optional low level warning reaction.