Centering the phase rings

The Zernike phase disc has five positions: 0 for brightfield observation (this position also has an iris diaphragm), 10, 20, 
40, and 100.

These positions correspond to the respective phase contrast objectives 10x, 20x, 40x and 100x.
When the condenser is in the “0” position, the objectives can be used for brightfield observation. For phase contrast, 
the condenser position should match the objective used, meaning that when the condenser is in position “40” the 
objective used should also be 40x.

• 

Rotate the 10× infinity phase contrast objective into the field of view, then set the Zernike phase disc to position 
“10” to match the objective 

• 

Take the eyepiece out of the tube and insert centering telescope in its place. When looking through the 
centering telescope, the dark and bright ring images should coincide with each other as shown in the pictures 
below. If the ring images cannot be observed clearly, focus the centering telescope first (A) and if this does not 
solve the issue then try to adjust the condenser by turning it up and down (B).

• 

If the bright ring and dark ring images do not coincide as shown below, adjust the position of the ring by 
moving the ring at the bottom of the condenser with the centering knobs (C). Move it until the bright and dark 
ring images superimpose. Repeat for all objectives/Zernike disc positions

 
 

 

 

 

 

 

15 

These positions correspond to the respective phase contrast objectives 10x, 20x, 40x and 100x. 
When the condenser is in the “0” position, the objectives can be used for brightfield observation. For phase 
contrast, the condenser position should match the objective used, meaning that when the condenser is in 
position “40” the objective used should also be 40x 
 
Rotate the 10× infinity phase contrast objective into the field of view, then set the Zernike phase disc to position 
“10” to match the objective   
 
Take the eyepiece out of the tube and insert centering telescope in its place. When looking through the centering 
telescope, the dark and bright ring images should coincide with each other as shown in the pictures below. If the 
ring images cannot be observed clearly, focus the centering telescope first (A) and if this does not solve the issue 
then try to adjust the condenser by turning it up and down (B). 
If the bright ring and dark ring images do not coincide as shown below, adjust the position of the ring by moving 
the ring at the bottom of the condenser with the centering knobs (C). Move it until the bright and dark ring 
images superimpose. Repeat for all objectives/Zernike disc positions 

 

 

 

 

 

 

 

 

 

 

 

              Not centered   

 

 

 

 

Centered properly 

 

 

 

 

 

  

Not centered

 
 

 

 

 

 

 

15 

These positions correspond to the respective phase contrast objectives 10x, 20x, 40x and 100x. 
When the condenser is in the “0” position, the objectives can be used for brightfield observation. For phase 
contrast, the condenser position should match the objective used, meaning that when the condenser is in 
position “40” the objective used should also be 40x 
 
Rotate the 10× infinity phase contrast objective into the field of view, then set the Zernike phase disc to position 
“10” to match the objective   
 
Take the eyepiece out of the tube and insert centering telescope in its place. When looking through the centering 
telescope, the dark and bright ring images should coincide with each other as shown in the pictures below. If the 
ring images cannot be observed clearly, focus the centering telescope first (A) and if this does not solve the issue 
then try to adjust the condenser by turning it up and down (B). 
If the bright ring and dark ring images do not coincide as shown below, adjust the position of the ring by moving 
the ring at the bottom of the condenser with the centering knobs (C). Move it until the bright and dark ring 
images superimpose. Repeat for all objectives/Zernike disc positions 

 

 

 

 

 

 

 

 

 

 

 

              Not centered   

 

 

 

 

Centered properly 

 

 

 

 

 

  

Centered properly

13

Содержание 78904-02

Страница 1: ...Installation and Operating Instructions MSU 400 Series Compound Microscopes Models 78904 02 04 07 10 13 78904 02 78904 07 78904 13...

Страница 2: ..._______________________________________________________ 11 Use of the S100x oil immersion objective_________________________________________________________ 12 Illumination____________________________...

Страница 3: ...comes in direct contact with the microscope parts In that case prevention of contamination lies in the handling of the slides as long as the slides are decontaminated before use and are undamaged and...

Страница 4: ...sed disinfection products for laboratories Before returning a microscope for repair or maintenance through a Cole Parmer dealer an RMA return authorization form together with a decontamination stateme...

Страница 5: ...e force when turning the knobs Make sure that the microscope system can dissipate its heat fire hazard Keep the microscope away from walls and obstructions for at least approximately 15 cm Never turn...

Страница 6: ...and S100x objectives are equipped with a spring mount to prevent damage to the front lens and the slide Model Type Objectives Working distance Stand Illumination Power Weight 78904 02 Binocular 4x 10...

Страница 7: ...er adjustment B Trinocular tube K Slide for polarization filter C Microscope head L Nosepiece D Transport handle M Objectives E Height adjustment condenser N Stage with X Y mechanical stage F Slide pr...

Страница 8: ...2 Install the objectives into the objective nosepiece from the lowest magnification to the highest in a clockwise direction from the rear side of the microscope When operating start using the low magn...

Страница 9: ...illumination For optimal contrast and resolution one should follow the below procedure 1 Place a specimen on the object stage and focus using the 4x objective with a fully opened iris diaphragm 2 Tur...

Страница 10: ...e field of view see image below This distance can be set by either pulling the tubes towards each other or pulling them away from each other This distance is different for each observer and thus shoul...

Страница 11: ...aphragm gradually 7 The condenser is centered correctly if the image remains in the center when you open the field diaphragm and inscribed to the field of view The field K hler diaphragm see A in pict...

Страница 12: ...the front lens doesn t touch the oil any longer Clean the front lens of the S100x objective 12 Always clean the front lens of the S100x objective with a piece of lens paper that is moistened with a dr...

Страница 13: ...te the 10 infinity phase contrast objective into the field 10 to match the objective Take the eyepiece out of the tube and insert centering telesc telescope the dark and bright ring images should coin...

Страница 14: ...reen F Fluorescence illuminator G Turret for LEDs Installation steps Take the microscope with fluorescence attachment fluorescence illuminator and head from its packaging Place the head on the fluores...

Страница 15: ...escence filters Blue Green Violet and Ultraviolet mounted in the turret fig 1 D and are indicated as B G V and UV on the turret There is also a brightfield option and it is indicated as 0 on the turre...

Страница 16: ...steps 1 Use objective 10x 2 Push in the field diaphragm switch the one in the front fig 2 A a light spot will be visible in the field of view 3 Adjust the centering if the light spot deviates from th...

Страница 17: ...Cleaning the optics When the eyepiece lens or front lens of the 10x or S40x objective are dirty they can be cleaned by wiping a piece of lens paper over the surface circular movements When this does...

Страница 18: ...roscopes come with a five 5 year factory warranty against manufacturing defects covering labor and parts Warranty does not apply for consumable accessories and parts such as bulbs batteries fuses cord...

Страница 19: ..._ ___________________________________________________________________________________________________ __________________________________________________________________________________________________...

Страница 20: ...2022 Cole Parmer Instrument Company LLC Toll free 1 800 323 4340 Phone 1 847 549 7600 Fax 1 847 247 2929 coleparmer com 1065Y62_M_CP_MSU 400_Series_Compound Microscopes_Manual...

Отзывы: