Disinfection and decontamination
•
Exterior casing and mechanical surfaces must be wiped with a clean cloth, dampened with a disinfectant
•
Soft plastic parts and rubber surfaces can be cleaned by gently wiping a clean cloth, dampened with a
disinfectant. Discoloration can occur if alcohol is used
•
The front lens of eyepieces and objectives are sensitive to chemicals. We recommend not to use aggressive
disinfectants but to use lens paper or a soft fiber-free tissue, damped in cleaning solution. Cotton swabs may
also be used. We recommend you use personal eyepieces without eyeshades in order to minimize risk
•
Never immerse or dip the eyepiece or objective into a disinfectant liquid! This will damage the component
•
Never use abrasive compounds or cleaners that may damage and scratch optical coatings
•
Properly clean and disinfect all possible contaminated surfaces of the microscope or contaminated accessories
before storing for future use. Disinfection procedures must be effective and appropriate
•
Leave the disinfectant on the surface for the required exposure time, as specified by the manufacturer. If the
disinfectant evaporates before the full exposure time, reapply disinfectant on the surface
•
Against bacteria, use a 70% aqueous solution of isopropanol (isopropyl alcohol) and apply for at least 30
seconds. Against viruses, we recommend to refer to specific alcohol or non-alcohol based disinfection products
for laboratories
Before returning a microscope for repair or maintenance through a Cole-Parmer dealer, an RMA (return authorization
form) together with a decontamination statement must be filled in! This document must be shipped together with the
microscope at all times
Handle with care
•
This product is a high quality optical instrument. Delicate handling is required
•
Avoid subjecting it to sudden shocks and impacts
•
Impacts, even small ones, can affect the precision of the instrument
Handling the LED
Note:
Always disconnect the power cord from your microscope before handling the LED bulb and power unit and
allow the system to cool down approximately 35 minutes to avoid burns
•
Never touch the LED with your bare hands
•
Dirt or fingerprints will reduce the life span and can result in uneven illumination, lowering the optical
performance
•
Use only original MSU-400 Series replacement LEDs
•
The use of other products may cause malfunctions and will void warranty
•
During use of the microscope the power unit will get hot; never touch it while in operation and allow the system
to cool down approximately 35 minutes to avoid burns
Dirt on the lenses
•
Dirt on or inside the optical components, such as eyepieces, lenses, etc., affects the image quality of your system
negatively
•
Always try to prevent your microscope from getting dirty by using the dust cover, prevent leaving fingerprints
on the lenses and clean the outer surface of the lens regularly
•
Cleaning optical components is a delicate matter. Please, read the cleaning instructions further on in this
manual