1. Place the power switch in the off position.
CAUTION: Power must be turned off before connecting the
external remote control cable to prevent damage to the drive.
2. Connect the cable from the external remote control to the mating
receptacle on the bottom panel.
3. Select type of remote control input and output required as follows:
a). Press and hold the MODE key while turning the power switch
to the “ON” (1) position. After two seconds, release the
MODE key. The initial display will show: “inP”. After two
seconds the display will show either 0–20 or 4–20.
NOTE:
Press the up (increment) or down (decrement) arrows to select
between 4–20 and 0–20 for current loop control.
b). Press the MODE key again. The initial display will show:
“out”. After two seconds the display will show either 0–20,
4–20, or 0–10.
NOTE:
Press the up (increment) or down (decrement) arrows to select
between 4–20 and 0–20 for current loop output, or 0–10 for voltage output.
4. Press the MODE key to select mode of operation. The LED’s indicate
the selected mode. Select either mA or V.
NOTE:
If only remote STOP/START, PRIME and/or CW/CCW is to be
used, the MODE control can be set to any of the three positions.
5. To adjust the voltage or current scaling for other than zero to full scale:
a). Press the MODE key and then the FLOW key at the same
time. The display will show “LO” and then the flow rate for
minimum current/voltage (factory default = 0).
b). Use the UP/DOWN arrow keys to change the flow rate for
minimum current/voltage.
c). Press the FLOW key. The display will show “HI” and then the
flow rate for maximum current/voltage. Use the UP/DOWN
arrow keys to change the flow rate for maximum
current/voltage or press the SIZE key to set it to maximum
flow rate (factory default). Press any other key to save and exit.
The same scaling will be used for both input and output. Each
tube size has its own scaling.
B/T
®
RAPID-LOAD
®
Peristaltic Pumps and Drive Operating Manual 3-5
Cole-Parmer
Section 3
Operation
Remote Control Setup