Command
:OUTPut
[n]:CHANnel
[m]:POWer:AVERagingtime?
Syntax
:OUTPut[n]:CHANnel [m]:POWer:AVERagingtime?<wsp>[MIN|MAX|DEF|SET|ALL]
Description
Query the averaging time for output power control
Parameters
MIN: Get the minimum programmable value
MAX: Get the maximum programmable value
DEF: Get the default programmable value
SET: Get the current set value
ALL: Returns all of the above parameters in a comma separated string
Response
Returns the string for the power averaging time of the specified channel
Example
OUTP1:CHAN1:POW:AVER? MIN -> 0.0000
Command
:OUTPut
[n]:CHANnel
[m]:POWer:NULLing
Syntax
:OUTPut[n]:CHANnel [m]:POWer:NULLing
Description
Starts dark power nulling. Use in conjunction with *OPC? to check for completion of dark
power nulling.
Parameters
None
Response
None
Example
OUTP1:CHAN1:POW:NULL
Command
:OUTPut
[n]:CHANnel
[m]:POWer:OFFSet
Syntax
:OUTPut[n]:CHANnel[m]:POWer:OFFSet<wsp><value>[DB|MDB]
Description
Set the user Output Offset reference for the specified channel
Parameters
<value>: Set to the user value with the specified units of dB (default) or mdB
Response
None
Example
OUTP1:CHAN1:POW:OFFS 5.0 DB
Command
:OUTPut
[n]:CHANnel
[m]:POWer:OFFSet?
Syntax
:OUTPut[n]:CHANnel [m]:POWer:OFFSet?
Description
Query the user Output Offset reference for the specified channel
Parameters
None
Response
Returns the string for the set user Output Offset reference for the specified channel
Example
OUTP1:CHAN1:POW:OFFS? -> 3.5000
Command
:INPut
[n]:CHANnel
[m]:AMODE
Syntax
:INPut[n]:CHANnel [m]:AMODE<wsp>[ABS|REL|OFFSET]
Description
Set the attenuation mode reference
Parameters
ABS: Set the attenuation reference to absolute
REL: Set the attenuation reference to relative
OFFSET: Set the attenuation reference to the user offset
Response
None
Example
OUTP1:CHAN1:AMODE ABS
Command
:INPut
[n]:CHANnel
[m]:AMODE?
Syntax
:INPut[n]:CHANnel [m]:AMODE?
Description
Query the attenuation mode reference for the specified channel
Parameters
None
Response
Returns a string (non-abreviated) for the current attenuation mode of the specified channel
Example
INP1:CHAN1:AMODE? -> ABSOLUTE