Modular Variable Optical Attenuator
|
VOA
Blade
Coherent Solutions
|
VOA
Blade (V1.4
)
41
13.5.2 Configuration Commands
Command
:CONTrol
[n]:CHANnel
[m]:MODE
Syntax
:CONTrol[n]:CHANnel [m]:MODE<wsp>[ATT|POW]
Description
Set the control mode for the specified channel
Parameters
ATT: Set to Attenuation control mode
POW: Set to Power control mode
Response
None
Example
CONT1:CHAN1:MODE ATT
Command
:CONTrol
[n]:CHANnel
[m]:MODE?
Syntax
:CONTrol[n]:CHANnel [m]:MODE?
Description
Query the control mode for the specified channel
Parameters
None
Response
Returns a string (non-abbreviated) for the control mode currently set for the specified
channel
Example
CONT1:CHAN1:MODE? -> ATTENUATION
Command
:OUTPut
[n]:CHANnel
[m]:POWer
Syntax
:OUTPut[n]:CHANnel [m]:POWer<wsp>[MIN|MAX|DEF|<value>[DBM|MDBM]]
Description
Set the constant power output for the specified channel
Parameters
MIN: Set to the minimum programmable value
MAX: Set to the maximum programmable value
DEF: Set to the default programmable value
<value>: Sets to the user value with the specified units (dBm is default)
Response
None
Example
OUTP1:CHAN1:POW 5.0
Command
:OUTPut
[n]:CHANnel
[m]:POWer?
Syntax
:OUTPut[n]:CHANnel [m]:POWer?<wsp>[MIN|MAX|DEF|SET|ALL]
Description
Query the power unit of measurement measured by internal power meter
Parameters
MIN: Get the minimum programmable value
MAX: Get the maximum programmable value
DEF: Get the default programmable value
SET: Get the desired set value
ALL: Returns all of the above parameters in a comma separated string
Response
Returns the string for the power output for the specified channel in dBm
Example
OUTP1:CHAN1:POW? ALL -> -45.0000,20.0000,10.0000,5.0000,4.9936
Command
:OUTPut
[n]:CHANnel
[m]:POWer:AVERagingtime
Syntax
:OUTPut[n]:CHANnel [m]:POWer:AVERagingtime
<wsp>[MIN|MAX|DEF|<value>[S|MS|US|NS]]
Description
Set the averaging time for output power control
Parameters
MIN: Set to the minimum programmable value
MAX: Set to the maximum programmable value
DEF: Set to the default programmable value
<value>: Sets the averaging time to the user value in the specified units (seconds are default)
Response
None
Example
OUTP1:CHAN1:POW:AVER 0.5 MS