2. Safety notes
Coolant
The Osscora surgical set is designed for use with
physiological saline solution. Use only suitable
irrigation fluids and comply with the medical data
and instructions from the manufacturer. Use the
supplied irrigation tubing set or accessories supplied
by Cochlear. Coolant bottle or coolant bag can be
obtained from a pharmacist.
Sterility of irrigation tubing set
Sterile irrigation tubing sets are supplied with the
equipment. These irrigation tubings are disposable
articles and must be thrown away after each
treatment!
Please note the expiry date and the relevant
regulations for disposal of irrigation tubing.
Rotational energy
Fast deceleration of the bur can, at times, cause the
selected torque to be overloaded as a result of the
rotational energy stored in the drive system.
Mains cable
Only connect to an earthed socket outlet.
Power failure
In the event of a power failure, if the control
unit is switched off, or when alternating
between programs, the last values set are
saved and reactivated on power-up.
System failure
A total system failure of the Osscora surgical
set does not constitute a critical fault.
Intermittent operating mode S3 (4min/10min)
When there is a current consumption of
max. 0.8 A for 91051 and 91052 and 1.6
A for 91053 (with pump switched on), the
permissible relative load time is 4 minutes to
10 minutes pause time.