14. Osscora contra-angle handpiece
Sterilisation and storage
Cochlear recommends sterilisation according to EN
13060 or EN 285, class B.
– Note the instructions of the steriliser manufacturer.
– Clean, disinfect, reassemble and lubricate before sterilising.
– Wrap the contra-angle handpiece and the accessory in sterile
goods packing according to EN 868-5.
– Make sure, that you only remove dry sterile goods.
– Store sterile goods dust-free and dry.
Approved sterilisation procedures
Follow your country-specific directives, standards and
guidelines.
– Steam sterilisation with sterilisers in accordance with
EN 13060 or EN 285, class B.
Sterilisation holding time: Minimum of 3 minutes at
134 °C (273.2 °F)
or
– Steam sterilisation with sterilisers in accordance
with EN 13060 or EN 285, class S.
The steriliser manufacturer must give their express approval
for the sterilisation of contra-angle handpieces.
Sterilisation holding time: Minimum of 3 minutes at
134 °C (273.2 °F).