16. Servicing
Regular checking of Osscora surgical set and
accessories
Regular servicing of function and safety of the Osscora
surgical set including the accessories is necessary
and should be carried out at least once every three
years, unless shorter intervals are prescribed by law.
The inspection must be undertaken by a qualified
organisation and needs to include the following
procedures:
– Visual inspection for outside damage.
– Measurement of enclosure leakage current.
– Measurement of patient leakage current.
– Visual inspection of internal components and electronic
boards on suspicion of safety interference, e. g. mechanical
damage of the enclosure or indicators for overheated or
burned (electronic) components.
We recommend that only authorised personnel undertake this
servicing and checking. For details, please contact your local
Cochlear office.