13. Disinfection, cleaning, sterilisation
Follow your country-specific directives, standards and
guidelines for cleaning, disinfection and sterilisation.
– Wear protective clothing.
– Clean and disinfect the motor
immediately after every treatment.
– Sterilise the motor following cleaning and
disinfection.
Control unit, foot control
Pre-disinfection
– If heavily soiled: Clean first with disinfectant cloths.
Only use disinfectants that have no protein-fixing
effects.
Manual cleaning and disinfection
The front panel of the control unit and the foot
control are sealed and may be wiped clean.
– Disinfection with surface disinfectants, wiping
disinfection is recommended.
– Use only surface disinfectants which do not contain chlorine
and which are certified by officially recognised institutes.
For USA:
Use EPA registered surface disinfectants.
– Note the manufacturer’s specifications for the use of the
surface disinfectants.