6. Electrical connections of the CleanMaster hood
Danger of electric shock!
Electricity supply to the control device
NOTE! Electrical connections may only be made by a professional electrician.
- The hood has to be electrified through the safety switch, which is visible near the hood and has
been named
- The control device’s switching power is max. 400 W
- The light is connected to the light group (connection cable as factory equipment)
Electricity supply connections:
Q1:2 (wire protection) - 230 VAC IN, 50 Hz, L
Q1:4 (wire protection) - 230 VAC IN, 50 Hz, N
X2: PE (left) 230 VAC IN, 50 Hz, PE
Electricity supply is brought to the casing through a strain
reliever. Note, the opened hatch may turn.
Electricity supply has to be connected to the hood/hood
group via a safety switch.
The safety switch has to be clearly visible and the label
included in the delivery must be attached to it!
!
CleanMaster
huuvan/ilmanvaihtokaton turvakytkin
hood/ventilation ceiling safety switch
Katkaise virta AINA ennen
huollon aloittamista.
before maintenance.