Appendix VI: Alarm-test procedure
(Authorized personnel only)
To test the alarm limits, the automatic fan control must be overruled, so the fan power can be set
by hand. The entire state of the cabinet will stay unchanged, except for the fan-power. The
required test-code is 7733.
In this state, each 30 second, two short audible signals will be heard, to be remembered at the
initial unsafe mode. After 30 minutes after the last button is touched, the fan returns to its work
mode.
Alarm-test procedure
1
Cabinet in work mode.
See
§ 5.3 Starting up the cabinet
on page
19.
•
Push button
(Menu).
•
Push button
(C) .
Cabinet in work mode
–
Safe
Monday 11:06:33
MENU
Socket=Off
A => Hour counter information
B => Status cabinet
C => Settings
A
B
C
Esc
2
Alarm test
•
Push 3x button
(>>).
•
Push button
.
•
Enter the test code: 7733.
•
Push button
.
Testing of alarms
A = Enter
>>
<<
Esc
Testing Alarms
0000 (Enter code)
A = Enter
>>
+
-
Esc
3
Alarm test settings
(NM = Normal mode,
PS = Power save mode,
DF = Data of cross Flow,
IF = Inflow, (not applicable)
CV = Current value)
•
Push button
(+) or
(-) in order to adjust the
fan speed. Value CV will change.
NM 500 Off 0 PS 300 CV 490
DF Off 0.00 PS 0.19 0.36 < 0.45 < 0.54
IF Off 0.00 PS 0.00 0.00 < 0.00 < 0.00
+
-
Esc