CLF
Laminar crossflow-cabinet
Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
22
5.7
Control UV (optional)
The UV-function is only visible if this option is installed. If not installed the function at the display
is invisible.
The UV-function may only be switched on with a completely closed workspace. See UV-Radiator
on page 16)
This prevents direct contact on the skin or eyes of the personnel.
Switching UV on
1
Make sure that the entire
workspace is closed.
Cabinet must be switched off or
in power safe mode.
The display shows “UV=Off”.
2
The UV will switch on.
The display shows “UV=On”.
•
Push button
(UV=Off).
Cabinet switched off
–
Not Safe
UV switched on
MENU
Socket=Off
UV=On
Switching UV off
3
The UV will switch off.
The display shows “UV=Off”.
•
Push button
(UV=On).
Cabinet switched off
–
Not Safe
UV switched off
MENU
Socket=Off
UV=Off
FL (Fluoresced lamp)-Lighting and UV cannot be switched on both at the same time. As a safety
precaution, one of them will automatically switch off.
5.8
Control sockets (Optional)
The sockets can always be switched on and off. If the cabinet is switched off, the sockets are
automatically switched off. Also if the cabinet is switched from work mode into power save mode,
the sockets are automatically switched off.
See technical specification (see
§9.2 General specifications on page 45
) for maximum voltage and
current (fuses).
Switching socket on
1
Cabinet in work mode.
See
§5.3 Starting up the cabinet
on page 19.
Cabinet in work mode - Safe
Monday 11:06:33
MENU
Socket=Off