CLF
Laminar crossflow-cabinet
Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
19
Function keys:
The function of keys 1, 2, 3 and 4 is variable. In the last line of the display the function is shown.
Signals:
Alarm: Confirmed alarm state
Alarm, flashing (along with acoustic signal):
Actual alarm
UV: Indication UV-light switched on
OK: Safe user mode
All equipment and, if possible, all materials required for working with the cabinet need to be
placed into the installation before starting up the cabinet. High equipment or objects can be placed
in the working space by sliding or hinging the window upwards. Be careful: keep the safety
regulations in consideration.
5.3
Starting up the cabinet
1
The lighting of the display will
light up.
The alarm indicator light is red
.
•
Connect the power cable.
Cabinet switched off
–
not safe
Monday 11:06:33
MENU
Socket=Off
2
The fan starts.
Cabinet is in start up mode.
The alarm indicator light is red
.
There is an acoustic alarm.
•
Push button
.
•
To acknowledge this alarm, push button
.
Starting to work mode
–
not safe
Monday 11:06:33
MENU
Socket=Off
3
If the cabinet is secured with a
user code, this code has to be
entered first.
The default code is 0001.
The user code can be changed
by the user, see Appendix IX:
User code on page 62.
•
Enter actual user code.
•
Confirm the code with button
.
User code:
0000 (Enter code)
A = Enter
>>
+
-
Esc
4
The light will switch on.
•
Push button
.
5
When the fan has reached the
right airspeed, the alarm
indicator will disappear.
The following display will be
shown.
However, the cabinet must operate for minimally 15
minutes more, before the cabinet is ready to work with.
Cabinet in work mode
–
safe
Monday 11:06:33
MENU
Socket=Off