CLF
Laminar crossflow-cabinet
Clean Air Techniek B.V.
UM-CLF-XX-EN05.doc
43
Figure 11 Positioning the left hand in the plastic bag
̇
Pull the pre-filter towel with your right hand;
̇
Put the filter into the plastic bag;
̇
Fold back the edge of the plastic bag, not touching the inner side of the plastic bag;
̇
Press the air out of the bag;
̇
Seal the plastic bag airtight;
̇
Remove the plastic bag and place it in an appropriate container;
̇
Put on new gloves;
̇
Place a new pre-filter towel;
̇
Remove the gloves and put them in the appropriate container.
8.3
Replacing the fluorescent-light (FL light)
When an FL light needs to be changed, switch off the power of the installation. The upper front
panel needs to be opened. Unscrew the nuts of the coverplate.The transparent light plate needs to
be removed in to the workarea. Replace the FL tube for a new one (see
Figure 12)
. Reposition all
parts on the installation.
DANGER FOR ELECTROCUTION
Figure 12 Position of the FL-tube