22
−
OPERATION
2
2.4
Standard setting mode
2.4.1
How to enter to standard setting mode
mL / min
<
Instantaneous flow rate display.
>
Press for 3 sec.
<
Switch output
>
Press once
<
Forcible output function
>
Press once
<
0 point adjustment
>
Press once
To Instantaneous flow rate display
To switch output
setting display
Press once
(
determination
)
To forcible output
setting display
Press once
(
determination
)
To 0 point adjustment
setting display.
Press once
(
determination
)
mL / min
To page 22
To page 24
To page 24
At instantaneous flow rate display, press
key for 3 sec. to enter to standard
setting mode.