background image

 CircuPool

 Help Guides:   

 

 

 

 

 

 

 

    

RJ Classic Series

 

13

 

 

2.

 

Low or Absent Water Flow through the Salt Cell

 

The water flow should always fully fill the entire Electrolytic Cell at all times. The  minimum flow requirement is typically 30 
gallons per minute. As the Electrolytic Cell has no ability to increase or decrease water flow, if the water is not completely 
filling the Cell then that indicates a problem with the pool's circulation system. Common things to check include: check the 
integrity  of  the  o-rings  on  the  pump  and  filter,  clean  or  backwash  your  filter,  check  the  water  pressure  from  your  pump 
motor, check for cracks or leaks in the pool's plumbing and plumbing unions that could be allowing air to be sucked into the 
circulation system.

 

3.

 

Bad Connection between Cell Cord and Cell Terminals

 

Ensure that all electrical connections are tight, free of corrosion, and fully seated. 

Models with 3-plug connectors 

 

Make sure the blue female cell cable connector is properly seated on the center brass post. 

 

Make sure that none of the prongs on each female terminal are broken off. 

 

Ensure that no dirt, debris or corrosion are causing a poor connection between the male and female terminals. If 
dulled by corrosion, steel wool or fine sandpaper can be used to remove the corrosion. 

 

Ensure that each female terminal is pressed fully on to each male terminal 

Models with single circular plug 

 

Use needle-nose pliers to crimp female connections on plug to ensure each has a tight connection to male terminals. 

 

Ensure that plug is 100% seated on to  male terminals. Removing the Cell's UV Shell case can allow easier access and 
show if the plug is pressed fully into place. 

 

Ensure that no dirt, debris or corrosion are causing a poor connection between the male and female terminals. If dulled 
by corrosion, steel wool or fine sandpaper can be used to remove the corrosion. 
 

4.

 

Cell has excessive scaling / mineral deposits / filter debris on titanium plates

 

The Electrolytic Cell should be free of calcium scale or any debris (hair, grass, etc..) that have accumulated between the titanium 
plates.  Frequency  of  cleaning  is  dependent  upon  your  pool's  water  chemistry,  but  for  many  people  is  required  just  a  couple 
times per season (see Saturation Index, page 5, for help balancing the water). If present, the Cell is not able to pass along power 
efficiently and needs to be cleaned. See page 18 for instructions.

 

5.

 

Loose or disconnected Titanium Plates in Electrolytic Cell caused by corrosive water

 

Remove  and  examine  the  titanium  plates  from  within  the  Cell.  Normal  wear  should  not  cause  any  significant  thinning  of  the 
metal, change of shape, degradation of edges, etching or pitting of surface, disconnection or separation of metal, or creation of 
any  holes  in  the  metal.  If  such  damage  is  present,  it  is  not  normal  and  is  important  to  identify  because  it  can  be  prevented 
through water chemistry maintenance. In order to disassemble the cell for easy inspection, see instructions on page 19. See the 
next few pages for examples.

 

The damage pictured in these examples is caused by the Saturation Index of the water being too low, causing the water to be 
corrosive. (see “Saturation Index” on page 5). The recommended range for the Saturation Index is (-0.2 to +0.2).

 

 

 

Содержание RJ Classic Series

Страница 1: ...22 How to Install the LCD Display Circuit Board pg 23 How to Install the Keypad pg 24 The complete RJ Classic manual is always available at Circupool com http www circupool com how to install a salt c...

Страница 2: ...consumed by sunlight and by breaking down organic material in your pool The level of chlorine you need to maintain depends on your CYA level and how much you use the pool It is important that you do...

Страница 3: ...calcium scaling You can raise TA with baking soda It is often best to make large TA adjustments in a couple of steps testing the water after each one as adding baking soda will also affect the pH and...

Страница 4: ...eliminate the accumulated phosphate is no longer necessary Phosphate removers can be added which allow the phosphates to be removed from the water A popular choice is Natural Chemistry s Phos Free wh...

Страница 5: ...less than 0 2 pool water is Corrosive Pool water may cause etching pitting dissolving and staining of walls grouting and plumbing It will also cause erosion to the titanium salt cell As Stabilizer In...

Страница 6: ...built in power meter that ensures that the full and complete amount of power is able to pass through the cell o The power meter is the bar graph that appears underneath the time and percentage output...

Страница 7: ...r chemistry report The Chlorine Stabilizer Cyanuric Acid or CYA level must be within range especially during the spring and summer months If there is not enough CYA in the pool then your chlorine will...

Страница 8: ...ess than is standard As a rule of thumb however run your system one hour for every ten degrees of ambient temperature in order to achieve both sufficient filtration and chlorination Periods of high us...

Страница 9: ...ith the titanium cell are excessive or have bridged the plates together This will cause the current across the cell to jump and activate the power protection Clean the cell plates See page 18 for inst...

Страница 10: ...anel from the black wall mounting bracket 2 Unscrew the six screws from the back of the control panel picture 1 below 3 Locate and unscrew the exterior inch hex nut located on the bottom the control u...

Страница 11: ...stomer requests 110V at the time of purchase Below are the internal wiring schematics for the CircuPool RJ Series If the CircuPool unit is wired for 220V and it is connected to a 110V power source all...

Страница 12: ...ystem does not appear to be affected by any cause listed here there may simply be environmental factors affecting the salinity reading and no action is required If Water Fault is illuminated the syste...

Страница 13: ...lug is 100 seated on to male terminals Removing the Cell s UV Shell case can allow easier access and show if the plug is pressed fully into place Ensure that no dirt debris or corrosion are causing a...

Страница 14: ...CircuPool Help Guides RJ Classic Series 14 Examples of GOOD titanium plates...

Страница 15: ...CircuPool Help Guides RJ Classic Series 15 Examples of DAMAGED titanium plates...

Страница 16: ...CircuPool Help Guides RJ Classic Series 16 Examples of DAMAGED titanium plates continued...

Страница 17: ...damaged there will be visible cracks in the black squares or black burn marks see picture below Once damaged the Water Fault light may illuminate or the unit may not power up at all B Damaged Capacito...

Страница 18: ...onnect the Cell Cord from the terminals on the Electrolytic Cell 3 Unscrew the two small threaded collars at the inlet and outlet side of the cell housing on bottom 4 Remove cell from the inlet and th...

Страница 19: ...side in between in the inlet and outlet pipes Model 2 ONLY ARROW A Method B Pull apart the UV shell by gripping the left and right sides near the cell connection Model 1 ONLY ARROW B Method C Use a sm...

Страница 20: ...Plates a Disconnect the cell connection from the male copper posts and unscrew the large threaded collar b Position a flathead screwdriver in the seam where the black cell connection plastic meets th...

Страница 21: ...he barrel unions a Unscrew the two small collars from the inlet and the outlet side of the cell housing b Carefully lift the cell housing off of the barrel unions If you are having trouble disconnecti...

Страница 22: ...both of the barrel unions Make sure the O ring is properly seated within the provided groove 5 YELLOW Optional Flat gasket that sits on top of the lip of both of the barrel unions If present make sure...

Страница 23: ...ing the bottom side in first and then press in the top making sure the circuit board is completely secured into the four clips Make sure the orientation is correct see picture above right 7 Reconnect...

Страница 24: ...six screws from the back of the control panel 3 Remove the front control panel cover 4 Disconnect the black green ribbon cable from the back of the display circuit board see picture below WARNING ALW...

Страница 25: ...the paper sticker away from the back of the new keypad 9 Place the green black ribbon cable through the provided groove and then press the new keypad onto the blue surface area 10 Reconnect the black...

Отзывы: