FIELD REPAIR/SERVICE
BLANKETROL III, Model 233
OPERATION AND TECHNICAL MANUAL
Page 92 of 131
B.
Drain the reservoir as described in Section (4-2.1.).
C.
Extend the front storage drawer as described in section (5-2.5).
D.
Locate the T-Shaped water filter assembly tucked under the water reservoir.
E.
Disconnect the assembly by loosening the two screw clamps (one on each side of
the water filter assembly).
F.
Remove the T-shaped water filter assembly.
G.
Insert the replacement water filter assembly so the cap and screen point down and
the arrow on the top points to the left. Connect one end to the hose from the water
flow indicator and the other end to the hose to the flow switch.
H.
Tighten both screw clamps.
I.
Refill the reservoir as described in Section (4-2.2.).
J.
Re-position the front storage drawer as described in section (5-2.6).
5-5.
REPLACEMENT OF THE PUMP HOUSING
A.
Obtain replacement pump housing.
B.
Drain the reservoir as described in Section (4-2.1.).
C.
Remove the rear enclosure panel as described in Section (5-2.1.).
D.
Locate the white pump housing assembly and the inlet and outlet hose connections
at the center left of the back of the unit as shown in Figure (5-1-D.).
CAUTION
Always drain the BLANKETROL III to a sanitary drain because
bio-contaminants
may be present in the unit’s water supply.
WARNING
Always
unplug the unit
before accessing internal components during service.
Failure to unplug the unit could result in electric shock.
CAUTION
Always drain the BLANKETROL III to a sanitary drain because
bio-contaminants
may be present in the unit’s water supply.