OPERATION
BLANKETROL III, Model 233
OPERATION AND TECHNICAL MANUAL
Page 35 of 131
AC.
Press the Up arrow or Down arrow to change the gradient variable offset to the
desired value.
1.
The microprocessor board beeps.
2.
The offset in the Status Display changes.
AD.
Press the GRADIENT VARIABLE button.
1.
The microprocessor board beeps.
2.
The LED in the corner of the button illuminates.
3.
The Patient display shows the patient's actual temperature.
4.
The WATER display shows the temperature in the water reservoir.
5.
The Status Display shows either:
* XXXXXX PATIENT
AUTO SETPT 37.0 C
Or Status Display shows:
PATIENT @SETPT
AUTO SETPT 37.0 C
* (”XXXXXX” represents “HEATING” or “COOLING”.)
The above depends upon the relationship of the patient temperature to the
Set Point temperature. Note that the Set Point may be displayed in
Fahrenheit.
6.
The pump is activated.
7.
The heater or compressor may also be activated.
8.
The water flow indicator on the right side panel begins to move. The water
moves from the unit to the blanket and returns to the unit.
9.
System operates in GRADIENT VARIABLE mode.
AE.
Press TEMP SET button
1.
The microprocessor board beeps.
2.
The LED in the corner of the button illuminates.
3.
The Status Display returns to set temperature mode.
AF.
Press the MANUAL CONTROL button.
1.
The microprocessor board beeps.
2.
The LED in the corner of the button illuminates.
3.
The WATER display shows the temperature of the water in the reservoir.
4.
The PATIENT display shows the temperature reading of the probe, as it is
held by the operator, if the probe reading is in the range of 10°C - 50°C (50°F
- 122°F).
5.
The Status Display shows either:
* XXXXXX WATER
MANUAL SETPT 37.0 C
Or Status Display shows:
WATER @SETPT
MANUAL SETPT 37.0 C
* (”XXXXXX” represents “HEATING” or “COOLING”.)