OPERATION
BLANKETROL III, Model 233
OPERATION AND TECHNICAL MANUAL
Page 29 of 131
9.
Push the male coupling until it SNAPS into position
10. Gently pull the connecting hose to assure a positive connection
I. Check that the blanket is laying flat and that the connecting hose to the unit is not
twisted or pinched.
J.
Check that the power switch of the unit is in “O” position (unit is off).
K.
Connect the plug to a properly grounded hospital grade receptacle.
2-3.2. Completing a System Test Routine
After arranging the equipment described in Section (2-3.1.), complete this System Test Routine
which describes what switches to press and the changes to observe.
NOTE:
If power is removed from the unit without actuating the power switch (I/O switch) a power fail
alarm is activated and the Power fail symbol LED located on the membrane control panel flashes.
To stop alarm, restore power to the unit and push the SILENCE ALARM button.
A.
Make sure that the power switch is in the “I” position (unit is turned on).
1. The switch illuminates green.
2.The microprocessor board goes through self-test.
3.The Status Display in the center of the membrane control panel flashes
CHECK SETPT.
If the above are not observed, consult the Troubleshooting Guide in Section (5-20.).
If they are observed, continue with the test routine.
B.
Press and hold the TEST INDICATOR button.
1. The alarm produces a beep
2. The LED's in the corner of the buttons and the low water and power
interruption LED’s flash on and then off.
3. All the displays flash.
The operator should note that all the displays and indicators illuminate. If they do not
illuminate, consult the Troubleshooting Guide in Section (5-20.). If they illuminate, continue
with the test routine. The Status Display continues to flash CHECK SETPT.
C.
Press the TEMP SET button.
1.
The microprocessor board beeps.
2.
The LED in the corner of the button illuminates.
3.
The Status Display shows:
SET TEMPERATURE
SETPT
37.0°C
D.
ENGLISH MEMBRANE ONLY:
Press the C/F button
.
1.
The SETPT temperature displayed in the Status Display changes from 37°C
to 98.6°F.
WARNING
Do not
by-pass ground lug (230V System).
Electrical hazards may result.