MAINTENANCE
BLANKETROL III, Model 233
OPERATION AND TECHNICAL MANUAL
Page 84 of 131
P.
Turn the BLANKETROL III off and remove the test jumper.
Q.
Allow the water temperature to fall below 42.0°C and turn the BLANKETROL III back
on.
R.
Confirm that the thermostatic disc has reset by verifying that EE02 is not shown in
the patient display.
S.
Press SILENCE ALARM button, TEMP SET button and increment button
simultaneously to reset the alarm.
Note: This step is required to reset the SNAP DISC REMOVE FROM SERVICE
alarm only.
4-10. TEMPERATURE ACCURACY CHECK
For temperature accuracy and verification we recommend using the CSZ TFRW 86171 Trimatic.
Follow directions enclosed in the Trimatic to test temperature accuracy.
4-11. DISPOSAL OF THE BLANKETROL II UNIT
Medical devices that have come in contact with patients contain the risk of bio-contamination.
This device generates no waste products or residues under normal use and normal cleaning
routines. Follow local State and Hospital guidelines regarding disposal of medical devices at the
end of their useful lives.