SPECIFICATIONS
BLANKETROL III, Model 233
OPERATION AND TECHNICAL MANUAL
Page 127 of 131
SECTION 7. SPECIFICATIONS AND CERTIFICATIONS OF THE
BLANKETROL III
BLANKETROL III MODEL 233 FEATURES
PHYSICAL
Dimensions:
17"W x 17"D x 37.5"H
(43.18cm.W x 43.18
cm. D x 95.25cm.H)
Weight:
Empty -131 lbs. (59.5kg)
Shipping
– 151 lbs. (68.5kg)
Cabinet Construction:
Powder coated steel with plastic top. Divided
reservoir. Bottom air discharge. Built-in handle.
Bumper guard. Storage Compartment. Two 4"
conductive, 360° swivel-
type casters and two 4”
non-conductive, 360° swivel-type locking
casters.
Ambient Temperature (during use):
15°C
– 30°C
(59°F
– 86°F)
Humidity (during use):
20% - 60%
Maximum Contact Surface Temperature:
41°C (105.8°F)
SAFETY SYSTEM
Maximum High Control Setting:
42.0°C (107.6°F)
High Limit Safety:
44.0°C
2.0°C (111.2°F
3.6°F)
High Limit Secondary Back-up Safety:
44.0°C
2.0°C (111.2°F
3.6°F)
Thermostatic Snap Disc:
46.0°C
2.0°C (114.8°F
3.6°F)
Minimum Low Control Setting:
4.0°C
(39.2°F )
Low Limit Safety:
2.0°C
2.0°C (35.6°F
3.6°F)
Low Limit Secondary Back-up Safety:
2.0°C
2.0°C (35.6°F
3.6°F)
Defective or Dislodged Probe Alarm:
Audible & Visual
Primary & Secondary High and
Secondary Low Limit Failure Alarm:
Audible & Visual
Low Water Alarm:
Audible & Visual
Defective Water Temp Sensor:
Audible & Visual
Water Flow Indicator:
Visual
Low Flow Alarm:
Audible & Visual
Water Temperature Sensor Fault Alarm:
Audible & Visual
THERMAL SYSTEM
Compressor:
1/3 HP
Heater:
800 Watts
NOTE:
Maximum expected heating capability
approximately 3°C (5.4°F) per minute.
Maximum expected cooling capability
approximately 5.5°C (9.9°F) per minute. Time
to heat from 23 ± 2°C to 37ºC is approximately
12 minutes with a blanket attached.
CIRCULATING SYSTEM
Divided Compartment Reservoir, 2 gallon (7.5
liters) total capacity. Error proof, quick-
disconnect fittings.
ELECTRICAL SYSTEM
Electrical Characteristics:
115V-127V±10%, 60Hz., 10.2 Amps
220V-230V±10%, 50Hz., 5.2 Amps
Power Cord:
14/3 SJT (115V),
Hospital grade plug
Standard European Cord (230V)
Leakage Current:
Under 300 µa 115
Under 500 µa 230
Circuit Breaker:
In Power Switch
Mains Isolation:
Two-Pole Mains Switch