OPERATION
BLANKETROL III, Model 233
OPERATION AND TECHNICAL MANUAL
Page 33 of 131
1.
The microprocessor board beeps.
2.
The LED in the corner of the button illuminates.
3.
The Status Display returns to set temperature mode.
U.
Press the Up arrow or the Down arrow so that the Set point displays 37°C (98.6°F).
Set the patient probe to a temperature reading between 30°C
– 43.5°C (86°F -
110.3°F).
V.
Press the AUTO CONTROL button.
1.
The microprocessor board beeps.
2.
The LED in the corner of the button illuminates.
3.
The WATER display shows the temperature of the water in the reservoir.
4.
The Status Display shows:
* XXXXXX PATIENT
AUTO SETPT 37.0 C
Or Status Display shows:
PATIENT @SETPT
AUTO SETPT 37.0 C
* (”XXXXXX” represents “HEATING” or “COOLING”.)
The above depends upon the relationship of the patient temperature to the
Set Point temperature. Note that the Set Point may be displayed in
Fahrenheit.
5.
The pump is activated.
6.
The heater or compressor may be activated.
7.
The water flow indicator on the right side panel begins to move. The water
moves from the unit through the blanket and returns to the unit.
8.
System operates in AUTO CONTROL mode.
W.
Press TEMP SET button
1.
The microprocessor board beeps.
2.
The LED in the corner of the button illuminates.
3.
The Status Display returns to set temperature mode.
X.
Press GRADIENT 10C button
1.
The microprocessor board beeps.
2.
The LED in the corner of the button illuminates.
3.
The WATER display shows the temperature of the water in the reservoir.
4.
The Status Display shows:
* XXXXXX PATIENT
AUTO SETPT 37.0 C
Or Status Display shows:
PATIENT @SETPT
AUTO SETPT 37.0 C
* (”XXXXXX” represents “HEATING” or “COOLING”.)