Digital Power Meter 66203/66204 User’s Manual
E-4
66203/66204 5A Range
Fluke 5520A (or
Fluke 5500A) Output
Parameter
Max. Spec.
Measured
Result
Min. Spec.
High Current
100V/4Arms, 60Hz
Irms 4.0090 3.9910
Low Current
100V/0.5Arms, 60Hz
Irms 505.50m 494.50m
66203/66204 2A Range
Fluke 5520A (or
Fluke 5500A) Output
Parameter
Max. Spec.
Measured
Result
Min. Spec.
High Current
100V/1.6Arms, 60Hz
Irms 1.6036 1.5964
Low Current
100V/0.2Arms, 60Hz
Irms 202.20m 197.80m
66203/66204 0.5A Range
Fluke 5520A (or
Fluke 5500A) Output
Parameter
Max. Spec.
Measured
Result
Min. Spec.
High Current
100V/0.4Arms, 60Hz
Irms 400.90m 399.10m
Low Current
100V/50mArms, 60Hz
Irms 50.550m 49.450m
66203/66204 0.2A Range
Fluke 5520A (or
Fluke 5500A) Output
Parameter
Max. Spec.
Measured
Result
Min. Spec.
High Current
100V/0.16Arms, 60Hz
Irms 160.36m 159.64m
Low Current
100V/20mArms, 60Hz
Irms 20.220m 19.780m
66203/66204 0.05A Range
Fluke 5520A (or
Fluke 5500A) Output
Parameter
Max. Spec.
Measured
Result
Min. Spec.
High Current
100V/40mArms, 60Hz
Irms 40.090m 39.910m
Low Current
100V/5mArms, 60Hz
Irms 5.0550m 4.9450m
66203/66204 0.02A Range
Fluke 5520A (or
Fluke 5500A) Output
Parameter
Max. Spec.
Measured
Result
Min. Spec.
High Current
100V/16mArms, 60Hz
Irms 16.036m 15.964m
Low Current
100V/2mArms, 60Hz
Irms 2.0220m 1.9780m
Содержание 66203
Страница 1: ...Digital Power Meter 66203 66204 User s Manual...
Страница 2: ......
Страница 3: ...Digital Power Meter 66203 66204 User s Manual Version 1 0 March 2014...
Страница 8: ...vi...
Страница 12: ......
Страница 22: ......
Страница 26: ......
Страница 76: ......
Страница 126: ......
Страница 130: ......
Страница 132: ......
Страница 136: ......
Страница 144: ......