Digital Power Meter 66203/66204 User’s Manual
5-44
[CALCulate:]COMParator:LIMit:ENEG
Type: Channel-Specific.
Description:
This command sets upper and lower boundaries of energy in GONG
mode.
Setting syntax: COMParator:LIMit:ENEG<space><NR2>,<NR2><PMT>
Setting parameters: <NR2>, -1 ~ 99999.9999, -1 denote don’t care.
Query syntax:
COMParator:LIMit:ENEG?[<space><MAX | MIN>]<PMT>
Return parameters: <NR2> , -1 ~ 99999.9999
Header on:
:COMPARATOR:LIMIT:ENEG<space><NR2>,<NR2><RMT>
Header off:
<NR2>,<NR2><RMT>
Separator 0: <NR2>,<NR2><RMT>
Separator 1: <NR2>;<NR2><RMT>
Example:
none
CALIBRATION Sub-system
CALibration:AUTO?
{<NR1>}
Type: Channel-Specific.
Description:
The command is used to force 662xx to re-calibrate the offset of
current. If ambient temperature variation is too much, user may
program this command to improve the accuracy.
Setting syntax: none
Setting parameters: none
Query syntax:
CALibration:AUTO?<PMT>,
CALibration:AUTO?<space><NR1><PMT>
Query parameters: <NR1>, 1 ~ 4: Channel 1 ~ Channel 4
Return parameters: <CRD>, OK | WAIT | FAIL
Header on:
:CALIBRATION:AUTO<space><CRD><RMT>
Header off:
<CRD><RMT>
Example:
none