Verification
10-7
4. Please confirm if the values of Meter, Load & DVM meet to that in
Table 10-7
Source Setting
Meter & Load
Measurement
V_Mon
Frequency
(Hz)
Voltage
(V)
V Specification
Specification
45/440
10V
9.64~10.36
0.179~0.221
350V
349.30~350.70
6.945~7.055
10.5 DC Performance Test
10.5.1 I monitor Measurement Verification
, switch AC Load to DC mode, the procedures are as follows.
1. Press
CONF./LOCAL
.
2. Press
1
and
ENTER
to enter SETUP
.
[ C O N F ]
C H O O S E = 1 1 . S E T U P
2 . R E M O T E
3 . C A L I B R A T I O N
4 . D I S P L A Y S E L E C T I O N
5 . F A C T O R Y D E F A U L T
6 . M O D E L I N F O R M A T I O N
7 . P A R A L L E L M O D E
3. Move the cursors of
▲
and
▼
to “1-1. MODE SELECT”
.
4. Press direction keys of
◄
,
►
or turn rotary knob to select “DC” and then press
ENTER
.
1 - 1 . M O D E S E L E C T = D C
/
A
C
[ S E T U P ]
1 - 2 . S H O R T = O F F
/ T O
G
G
L
E
/
H
O
L
D
1 - 3 . L O C K M O D E = R O T A R Y
/ K E Y P A
D
/
A
L
L
1 - 4 . S O U N D = O N
/
O
F
F
1 - 5 . C F / P F S E T T I N G = B O T
H
/ C F
/
P
F
1 - 6 . T I M I N G M O D E = O F F
/
H O L D U P /
T
R
A
N
S
F
E
R
1 - 7 . E X T E R N A L = O F F
/ O N
1 - 8 . A B A M O D E
= E N A B L E
/
D I S A B L E
Содержание 63800 Series
Страница 1: ......
Страница 2: ......
Страница 3: ...Programmable AC DC Electronic Load 63800 Series Operation Programming Manual Version 2 0 July 2017 ...
Страница 8: ...vi ...
Страница 28: ......
Страница 148: ......
Страница 174: ......
Страница 180: ......
Страница 192: ......
Страница 194: ......
Страница 196: ......