Local Operation
3-23
conductance : 0.01+(0.01
×
0.5%+1
×
0.5%)
to
0.01
−
(0.01
×
0.5%+1
×
0.5%)
resistance
: 66.445
Ω
to
202.02
Ω
To set programmed CR mode is by pressing
CR
. Users can press the arrow key to select the
setting parameters on the LCD display, including resistor value and I
rms
(max).
R s e t = 5 . 0 0 I r m s ( m a x ) = 5 0 . 0
U N I T :
Ω
, A
V = 0 . 0 0 P = 0 . 0 0
I = 0 . 0 0 S = 0 . 0 0
A C
The procedures of setting
R
set
= 10.50
and
I
rms
(max) = 20.0
, are described as below:
1. Move the cursor to the command line of “R
set
=”.
R s e t = 5 . 0 0 I r m s ( m a x ) = 5 0 . 0
U N I T :
Ω
, A
2. Press
1
,
0
,
.
,
5
then press
ENTER
to change the value to “10.50”.
R s e t = 1 0 . 5 0 I r m s ( m a x ) = 5 0 . 0
U N I T :
Ω
, A
3. Move the cursor by pressing (
◄
,
►
) to the command line of “I
rms
(max)=”.
R s e t = 1 0 . 5 0 I r m s ( m a x ) = 5 0 . 0
U N I T :
Ω
, A
4. Press
2
,
0
then press
ENTER
to change the value to “20.0”.
R s e t = 1 0 . 5 0 I r m s ( m a x ) = 2 0 . 0
U N I T :
Ω
, A
5. After setting the parameter of the load, the
LOAD ON/OFF
key should be pressed to turn
on the load, so that the load can sink the current. (For the detailed information please
refers
section 3.4.5
If there is error during the procedure of inputting value, press
CLEAR
to clear
at any time and then input again.
Содержание 63800 Series
Страница 1: ......
Страница 2: ......
Страница 3: ...Programmable AC DC Electronic Load 63800 Series Operation Programming Manual Version 2 0 July 2017 ...
Страница 8: ...vi ...
Страница 28: ......
Страница 148: ......
Страница 174: ......
Страница 180: ......
Страница 192: ......
Страница 194: ......
Страница 196: ......