Freeze-dryer Epsilon 1-4 LSCplus
Freeze-dryer Epsilon 2-4 LSCplus
6 Operation
Version 07/2022, Rev. 3.0 of 20/07/2022
• sb
87 / 152
Translation of the original operating manual
After the completion of a phase, the system switches to the next phase
without switching the unit to standby. The transition from "freezing" to
"warm-up" takes place automatically. After the completion of the warm-up
phase, this dialog box will be displayed:
Fig. 56: Dialog box after the completion of the warm-up phase
The unit will remain in the warm-up phase until you confirm.
The transition from "main drying" to "final drying" again takes place
automatically. After the completion of the "final drying" phase, the system
displays another enquiry with which the freeze-drying process will be
terminated. The unit remains in the "run" mode until the enquiry is
confirmed.
The process can be stopped any time by pressing the "Stop" button. In this
case, the unit will be switched to standby.
Pos : 189 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 190 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/060 Betrieb/LSC plus /060- 0050-0022- 0011 Sollwert eingaben im manuell en M odus Epsil on @ 72\mod_1593154334624_68. docx @ 766851 @ 4 @ 1
6.6.2.1
Entering set values in the manual mode
The system has stored set values for every phase, and for every value
there are pre-defined value ranges that can be determined in the various
dialog boxes by pressing the buttons "min" and "max".
In order to protect the product, a
→
safety pressure
value can be entered in
every drying section.