Freeze-dryer Epsilon 1-4 LSCplus
Freeze-dryer Epsilon 2-4 LSCplus
6 Operation
Version 07/2022, Rev. 3.0 of 20/07/2022
• sb
103 / 152
Translation of the original operating manual
Fig. 71: The freeze-
drying process can be started with the aid of the button “Program start”.
• During the freeze-drying process, the description of the active program
can also be called up from the main window "Program" and via the
button "Info".
• When the "Stop" button is pressed, the freeze-dryer switches to the
standby
– manual freezing mode.
NOTE
The program starts with section 02 "Freezing" by default. If, however,
section 01 "Load" is to be used as the starting point of the program, this
section must be selected manually. In this case, the button "Program
continue" will be displayed after the start of the program. When the
precooling (loading) process is complete, this button must be pressed in
order to continue with the program.
Pos : 201 /010 Univ ers almodul e/ Leerz eile @ 0\mod_1202116244500_0.docx @ 114 @ @ 1
Pos : 202 /200 Chris t/360 GT-BA Labor-Pil ot ( STAND ARDMO DULE)/060 Betrieb/LSC plus /060- 0050-0033- 0050 Programm lösc hen @ 72\m od_1593154342016_68. docx @ 767003 @ 4 @ 1
6.6.3.5
Deleting a program
The PGMplus programmer module offers 32 program storage locations. If
they are all occupied, a program must be deleted before a new one can be
created.
• Select the program to be deleted from the list in the main window
"Program".
• Press the button with the program name in order to call up a window
with the program data.
• Press the button “Delete program”. The system will then display a dialog
box with an enquiry.
• Following the confirmation of the enquiry, the program will be deleted.
The program storage location on the list is now empty and the number is
not shown on the list.
Pos : 203 /010 Univ ers almodul e/ Seit enwechs el @ 0\mod_1202116244312_0.docx @ 105 @ @ 1